Falsely low urinary albumin concentrations after prolonged frozen storage of urine samples.

Brinkman JW, de Zeeuw D, Duker JJ, Gansevoort RT, Kema IP, Hillege HL, de Jong PE, Bakker SJ

Publication: Clin Chem (2005) • Vol 51 • Page 2181-3PMID: 16244297View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage and mixing on urinary albumin concentration (UAC) in specimens with initial UACs in the normo- and microalbuminuric ranges.

Conclusion

Significant decreases in UAC were seen when specimens were stored at -20 degrees C. Specimen mixing by vortex or hand inversion was able to attenuate these decreases for shorter storage periods. Notably, specimens with high UACs (200-500 mg/L) showed a significantly attenuated decrease in UAC after frozen storage compared to specimens falling in the lower UAC categories, and this was true regardless of duration of storage.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage and mixing on UAC in specimens with initial UACs in the normo- and microalbuminuric ranges. All specimens were collected and held at 7 degrees C until freezing or analysis. Specimens that were frozen at -20 degrees C were thawed at 7 degrees C for 1-3 days prior to analysis.

Summary of Findings

Specimens that were not mixed after frozen storage for 3-5 months showed significant decreases (-28.6%) in UAC compared to baseline values measured in aliquots held at 7 degrees C (p<0.001). Specimens that were mixed by either vortex or hand inversion after 3-5 months of frozen storage did not show significant changes in UAC. However, specimens that were frozen for 18-24 months and mixed prior to analysis showed significant decreases in UAC after both hand inversions (-26.2%) and vortexing (-24.7%) (p<0.001 for both), but no significant differences between mixing methods were observed, regardless of storage duration or UAC concentration category. Specimens with high UACs (200-500 mg/L) showed a significantly attenuated decrease in UAC after frozen storage compared to specimens falling in the lower UAC categories (p<0.001), and this was true regardless of duration of storage.

  • Biospecimens

    FluidUrine
  • Preservation Types

    FrozenOther Preservative
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    0-10 mg/L UAC
    10-20 mg/L UAC
    20-40 mg/L UAC
    200-500 mg/L UAC
    40-80 mg/L UAC
    80-200 mg/L UAC
    PreaquisitionDiagnosis/ patient condition
    Microalbuminuria
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    Hand inversion
    None
    Vortexing
    Biospecimen PreservationType of fixation/preservation
    Frozen
    Refrigeration
    StorageStorage duration
    12-18 months
    18-24 months
    3-5 months
    5-8 months
    8-12 months
    StorageStorage temperature
    -20 degrees C