Effect of in vitro hemolysis on chemical values for serum.

Frank JJ, Bermes EW, Bickel MJ, Watkins BF

Publication: Clin Chem (1978) • Vol 24 • Page 1966-70PMID: 709831View on PubMed

Purpose

The purpose of this paper was to determine the effects of hemolysis on clinical chemistry analytes in serum.

Conclusion

Increasing hemolysate concentration in serum increased potassium, iron, aspartate aminotransferase, lactate dehydrogenase (LDH), alpha-hydroxybutyrate dehydrogenase, creatinine kinase (CK) and acid phosphatase and decreased bicarbonate and bilirubin levels. The magnitude of the changes in iron, CK and acid phosphatase levels with increasing hemolysate were dependent on measurement method.

Studies (1)

Purpose

The purpose of this study was to determine the effects of hemolysate concentration on measured clinical chemistry analytes in serum measured by several methods. Erythrocytes obtained from EDTA-whole blood and lysed by the addition of de-ionized water and sodium saponin were added to thawed pools of serum. Serum containing hemolysate was preserved with disodium citrate monohydrate and stored at -20 degrees C for less than 24 h before analysis for all analytes except acid phosphatases, which were assayed within 48 h.

Summary of Findings

For all measurement methods, increasing hemolysate concentration resulted in increases in aspartate aminotransferase (66-84%), LDH (137-155%), alpha-hydroxybutyrate dehydrogenase (153%), CK (35-130%), acid phosphatase (15-315%), potassium (16-19%) and iron (10-24%) and decreases in bilirubin (-28%) and bicarbonate (-7.5%) levels . However, uric acid, urea nitrogen, albumin, alkaline phosphatase, cholesterol, total protein, sodium, magnesium, calcium, inorganic phosphorous and chloride were unaffected by hemolysate concentration. With the exception of CK,acid phosphatase and iron, the magnitude of changes observed was similar among methods. The magnitude of the increase in CK with the addition of 2800 mg/L hemoglobin was much greater when measured using the DuPont aca (130%) than when measured using the Gilford 3500 with Worthington reagents (35%). The magnitude of the increase in acid phosphatase levels with the addition of 2800 mg/L hemoglobin was much greater when measured using Bessey, Lowry and Brock method (315%) than when measured using the Bodansky method (15%) or the DuPont aca (22%). Similarly, while only a 10% increase in iron with the addition of 2800 mg/L hemoglobin was identified by the DuPont aca method a 24% increase was identified using the Henry method.
  • Biospecimens

    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    Electrolyte/MetalIon selective electrode
    Electrolyte/MetalFlame emission photometry
    Electrolyte/MetalAtomic absorption spectroscopy
    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    1440 mg hemoglobin/L
    320 mg hemoglobin/L
    Less than 90 mg hemoglobin/L
    More than 2800 mg hemoglobin/L
    Biospecimen Aliquots and ComponentsHemolysis
    Chemically-induced
    Hemolysate added
    No hemolysate added
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Atomic absorption spectrophotometer
    Bessey, Lowry and Brock method
    Bodansky
    Continuous flow analysis on Technicon SMA 12/60
    DuPont aca
    Flame photometer
    Gilford 3500 using Dade reagents
    Gilford 3500 using Worthington reagents
    Henry method
    Ion selective electrode