Agreement between paired blood gas values in samples transported either by a pneumatic system or by human courier.

Victor Peter J, Patole S, Fleming JJ, Selvakumar R, Graham PL

Publication: Clin Chem Lab Med (2011) • Vol 49 • Page 1303-9PMID: 21619479View on PubMed

Purpose

The purpose of this paper was to determine the effects of transportation through a pneumatic tube system (PTS) on arterial blood gas values.

Conclusion

High levels of agreement were seen for pH, bicarbonate, partial pressure of carbon dioxide (PaCO2), and oxygen saturation levels between specimens that were transported by manual courier or PTS. For specimens with partial pressure of oxygen (PaO2) values <160 mM Mercury (Hg), transportation by PTS resulted in falsely high PaO2 values. On the other hand, for specimens with PaO2 values >160 mM Hg, transportation by PTS resulted in falsely low PaO2 values when transportation by manual courier was taken as the standard.

Studies (1)

Purpose

The purpose of this study was to determine the effects of transportation by PTS versus manual courier on arterial blood pH, PaCO2, PaO2, bicarbonate, and oxygen saturation of 50 paired specimens. Prior to transport, samples were visually checked to ensure there were no air bubbles. Once transport was initiated, specimens were in the PTS for 40-80 s while transport by human courier took approximately 5 min. Transport by either method involved delays due to batching of specimens, and in some cases, specimens were collected and stored in ice.

Summary of Findings

High levels of agreement were seen for pH, bicarbonate, PaCO2, and oxygen saturation levels between specimens that were transported by manual courier or PTS. Despite the high correlation (0.98) and small average difference between PaO2 for pneumatic and manually transported specimens (-0.9 mM Hg), differences among individual pairs of specimens were larger than 40 mM Hg. For specimens with PaO2 values <160 mM Hg, transportation by PTS resulted in falsely high PaO2 values, but for specimens with PaO2 values >160 mM Hg, transportation by PTS resulted in falsely low PaO2 values when transportation by manual courier was taken as the standard. The differences in time between sampling and analysis for transportation by manual courier (average 35.7 +/- 23.2 min, range 10-100 min) and transportation by PTS (average 38.6 +/- 22.1 min, range 10-98 min) did not affect PaO2, but instead the authors conclude that exaggerated movement of oxygen between the specimen and the air during PTS transportation may explain the altered PaO2 values.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    PaO2 <160 mM Hg
    PaO2 >160 mM Hg
    StorageSpecimen transport duration/condition
    35.7 +/- 23.2 min (manual transportation)
    38.6 +/- 22.1 min (PTS transportation)
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system