Evaluation of lyophilized heparin syringes for the collection of arterial blood for acid base analysis.
Crockett AJ, McIntyre E, Ruffin R, Alpers JH
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of lyophilized heparin on PO2, PCO2 and pH in different volume aliquots of venous and arterial blood from a healthy volunteer. Venous blood anticoagulated with sodium heparin was collected from a single healthy volunteer on 5 different days. After gas and pH levels were adjusted with a tonometer, aliquots were transferred to plain glass syringes or Omnistik syringes containing lyophilized heparin. Arterial blood was collected using a 3 way stop-cock into a plain glass tube containing liquid sodium heparin or an Omnistik syringe. Specimens were stored on ice for up to 25 min prior to analysis.
Summary of Findings
PO2, pCO2 and pH were not affected by collection of 1 mL or 2 mL of blood in Omnistik syringes containing lyophilized heparin or in glass syringes, nor was there an effect on the range of values.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
GasClinical chemistry/auto analyzerSmall moleculepHPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anatomical location of blood draw ArteryVeinBiospecimen Acquisition Anticoagulant Lyophilized heparinNoneSodium heparinBiospecimen Aliquots and Components Aliquot size/volume 1 mL2 mL