Simultaneously obtained skin-puncture serum, skin-puncture plasma, and venous serum compared, and effects of warming the skin before puncture.
Blumenfeld TA, Hertelendy WG, Ford SH
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection method, including hand warming prior to skin puncture, and the use of serum rather than ammonium heparin plasma on the measurement of 15 clinical chemistry analytes in blood from healthy individuals. Skin temperature was increased by applying warmed surgical dressings to the hands for 3 min prior to skin puncture. Serum was collected into serum separator tubes (SST).
Summary of Findings
Lactate dehydrogenase (LDH), alkaline phosphatase, urea nitrogen, calcium, chloride, free hemoglobin, glucose and potassium levels were similar in plasma and serum collected through skin puncture. However, significant but clinically irrelevant depressions in magnesium and sodium as well as elevations in phosphorous levels were observed in serum compared to plasma collected through skin puncture. Lower levels of potassium, calcium and total protein that were clinically and statistically significant were observed in plasma and serum collected through skin puncture compared to plasma collected through venipuncture. Clinically relevant but statistically insignificant increases in glucose were found in plasma and serum collected through skin puncture compared to plasma collected by venipuncture. Free hemoglobin was significantly lower when specimens were collected by venipuncture than by skin puncture, but the difference was judged to be clinically irrelevant. There were no differences in levels of LDH, alkaline phosphatase, urea nitrogen, chloride, magnesium, and phosphorous between specimens collected by skin puncture and venipuncture. Warming of the hands to 37-37.5 degrees C from 34-35 degrees C did not affect the levels of any of the clinical chemistry analytes in serum or plasma collected by skin puncture.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerElectrolyte/MetalFlame emission photometryProteinClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerProteinSpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anatomical location of blood draw CapillaryVeinBiospecimen Acquisition Method of fluid acquisition Finger/heel prick samplingFinger/heel prick with warmingVenipunctureBiospecimen Aliquots and Components Blood and blood products PlasmaSerum