The purpose of this study was to determine the effects of collection method and container on the levels of clinical chemistry analytes in serum. Skin puncture blood was collected into a microtainer, while venipuncture blood was collected into a vacutainer.
Summary of Findings
Sodium, calcium, phosphorous, creatinine, and total protein were lower in serum collected by skin puncture than by venipuncture. In contrast, potassium, chloride, and TSH were higher in serum collected by skin puncture instead of venipuncture. Collection method had no effects on urea, bilirubin, ALAST, ASAT, LD, insulin or thyroxin. Growth hormone levels were unaffected by collection method in 3 of 9 patients and were below the limit of detection in the remaining 6. When blood was collected by venipuncture and then transferred to a microtainer, sodium and calcium levels were decreased and TSH was increased compared to the aliquot of the specimen kept in the vacutainer, indicating the effects for these three analytes were tube dependent, not collection method dependent.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerGlycoproteinRadioimmunoassaySmall moleculeClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerProteinRadioimmunoassayPeptideRadioimmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anatomical location of blood draw CapillaryVeinBiospecimen Acquisition Method of fluid acquisition Finger/heel prick samplingVenipunctureBiospecimen Acquisition Type of collection container/solution MicrotainerVacutainer