Superheating antigen retrieval.
Leong AS, Lee ES, Yin H, Kear M, Haffajee Z, Pepperall D
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of different antigen retrieval techniques on immunostaining using 42 routine diagnostic antibodies and multiple postmortem and surgical tissue specimens. Most antigen retrieval was performed in citrate buffer, but for certain antigens, ethlenediamine tetraacetic acid (EDTA) or tris buffer were used.
Summary of Findings
For all antibodies except anti-cytokeratins AE1/3, 34BE12, Cam 5.2, and anti-bovine keratin, superheating to 120 degrees C yielded immunostaining that was superior to other methods of antigen retrieval. For these 4 exceptions, microwaving in a pressure cooker or at 98 degrees C was gave slightly better immunostaining results. Ultrasound mediated antigen retrieval was inferior to microwave mediated antigen retrieval and only yielded slightly improved immunostaining compared to when no antigen retrieval technique was employed. Duration and wattage had no effects. Antigen retrieval through heating a pressure cooker by microwaving was comparable to microwaving at 98 degrees C.
Biospecimens
TissueHeartTissueSpleenTissueKidneyTissueLiverTissuePancreasTissueColorectalTissueStomachTissueBreastTissueSalivary GlandTissueEsophagusTissueTongueTissueTonsilTissueProstateTissueUterusTissueBladderTissueLungTissueSkinPreservation Types
FormalinDiagnoses
NeoplasticBenignNeoplasticMelanomaAutopsyNormalNeoplasticLymphomaNeoplasticCarcinomaPlatforms
ProteinImmunohistochemistryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Analyte Extraction and Purification Antigen retrieval Citrate buffer, pH 6.0EDTA, pH 8.0Microwaving at 120 degrees C, 1.9 bar pressure for 5 minMicrowaving at 98 degrees C for 10 minNo antigen retrievalPressure cooker heated by microwave for 5 minTris buffer, pH 10.0Ultrasound at 40 watts for 100 sUltrasound at 40 watts for 40 sUltrasound at 70 watts for 100 sUltrasound at 70 watts for 40 s