Skip to main content

Differential gene expression profiles are dependent upon method of peripheral blood collection and RNA isolation.

Asare AL, Kolchinsky SA, Gao Z, Wang R, Raddassi K, Bourcier K, Seyfert-Margolis V

Publication: BMC Genomics (2008) • Vol 9 • Page 474PMID: 18847473View on PubMed

Purpose

The purpose of this paper was to determine the effects of RNA extraction method on gene expression analysis from stimulated and unstimulated blood using microarrays.

Conclusion

RNA extraction using the Tempus system resulted in a higher average RNA yield, ratio of absorbance at 260 to 280, and percentage of present calls than RNA extraction using the PAXgene kit, and lower 3' to 5' GAPDH ratio than PAXgene. Hierarchical clustering properly clustered specimens based on phytohemagglutinin (PHA) stimulation, but a secondary clustering based on extraction method was also observed. A number of immune transcripts were confirmed to be differentially expressed after PHA stimulation using PCR with MALDI-TOF mass spectrometry, but for PAXgene extracted specimens, variability precluded statistical significance. Collection of specimens into lithium heparin tubes resulted in changes in the expression of B cell translocation transcript.

Studies (1)

Purpose

The purpose of this study was to determine the effects of blood preservation method and RNA extraction method on gene expression analysis from PHA stimulated and unstimulated blood using microarrays. Unless otherwise specified blood was collected into tubes without anticoagulant. Blood was transferred to the PAXgene or Tempus tubes.

Summary of Findings

RNA extraction using the Tempus system resulted in a higher average RNA yield, ratio of absorbance at 260 to 280, and percentage of present calls than extraction using the PAXgene kit, and lower 3' to 5' GAPDH ratio than PAXgene. Hierarchical clustering properly clustered specimens based on PHA stimulation, but a secondary clustering based on extraction method was also observed. 538 genes were upregulated, and 392 genes were downregulated by PHA stimulation when RNA was extracted with Tempus, but when RNA was isolated with PAXgene, 400 genes were upregulated, and 539 were downregulated by PHA stimulation. 56% of the upregulated genes and 40% of the downregulated genes showed similar directional changes for both extraction kits. A number of immune transcripts were confirmed to be differentially expressed after PHA stimulation using PCR with MALDI-TOF mass spectrometry, but in PAXgene extracted specimens, variability precluded statistical significance. Collection of specimens into lithium heparin tubes resulted in changes in the expression of B cell translocation transcript compared to collection in tubes without anticoagulant.

  • Biospecimens

    FluidBlood
  • Preservation Types

    PAXgeneOther Preservative
  • Diagnoses

    Not specified
  • Platforms

    RNADNA microarray
    RNAMALDI-TOF MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    None
    Biospecimen Aliquots and ComponentsBiospecimen components
    0 ug/ml PHA
    25 ug/ml PHA
    Biospecimen PreservationType of fixation/preservation
    PAXgene
    Tempus
    Analyte Extraction and PurificationAnalyte isolation method
    PAXgene
    Tempus