Evaluation of an improved blood-conserving POCT sampling system.
Weiss M, Dullenkopf A, Moehrlen U
Purpose
The purpose of this paper was to determine the effects of blood collection method on hematological and clinical chemistry analytes.
Conclusion
There were no clinically relevant effects of sampling method on the measurement of any analytes, but potassium levels were elevated to a level which was statistically but not clinically significant when analyzed. However, this was only true for one of two similar iSTAT machines tested.
Studies (1)
Purpose
The purpose of this study was to compare measurement of clinical chemistry and hematological analytes in blood collected directly into modified iSTAT sampling cartridges and collected in a syringe and transferred to an iSTAT cartridge. Hematocrit, sodium, potassium, calcium, glucose, bicarbonate, base excess, pH, hemoglobin, oxygen saturation, and partial pressure carbon dioxide and oxygen were measured using the iSTAT CG8+ cartridge on one of two similar iSTAT devices.
Summary of Findings
There were no clinically relevant effects of sampling method on the measurement of any analytes, but potassium levels were elevated to a level which was statistically but not clinically significant when analyzed. However, this was only true for one of the two similar iSTAT machines tested.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerCell count/volumeHematology/ auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Blood-conserving sampling lineSyringe draw