Evaluation of an improved blood-conserving POCT sampling system.

Weiss M, Dullenkopf A, Moehrlen U

Publication: Clin Biochem (2004) • Vol 37 • Page 977-84PMID: 15498525View on PubMed

Purpose

The purpose of this paper was to determine the effects of blood collection method on hematological and clinical chemistry analytes.

Conclusion

There were no clinically relevant effects of sampling method on the measurement of any analytes, but potassium levels were elevated to a level which was statistically but not clinically significant when analyzed. However, this was only true for one of two similar iSTAT machines tested.

Studies (1)

Purpose

The purpose of this study was to compare measurement of clinical chemistry and hematological analytes in blood collected directly into modified iSTAT sampling cartridges and collected in a syringe and transferred to an iSTAT cartridge. Hematocrit, sodium, potassium, calcium, glucose, bicarbonate, base excess, pH, hemoglobin, oxygen saturation, and partial pressure carbon dioxide and oxygen were measured using the iSTAT CG8+ cartridge on one of two similar iSTAT devices.

Summary of Findings

There were no clinically relevant effects of sampling method on the measurement of any analytes, but potassium levels were elevated to a level which was statistically but not clinically significant when analyzed. However, this was only true for one of the two similar iSTAT machines tested.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionMethod of fluid acquisition
    Blood-conserving sampling line
    Syringe draw