Perchlorate (Irenat®) may falsely lower measured ionised calcium.
Gruber M, Nehring C, Creutzenberg M, Graf B, Hopf S
Purpose
The purpose of this paper was to compare native and spiked ionized calcium measurements between 5 blood gas analyzers after blood specimens were spiked with perchlorate.
Conclusion
The measurement of ionized calcium was falsely decreased in specimens containing perchlorate when measured by the Rapidlab 865 and 405 blood gas analyzers, as well as the ABL 815 flex, but not the GEM 4000 and cobas b221 blood gas analyzers.
Studies (1)
Purpose
The purpose of this study was to compare ionized calcium measurements between 5 blood gas analyzers in blood spiked with calcium chloride and/or different concentrations of perchlorate. In addition, the stability of perchlorate in blood stored at room temperature was examined.
Summary of Findings
The measurement of native ionized calcium concentration was unaffected by the addition of perchlorate to blood specimens using the GEM 4000 and cobas b221 blood gas analyzers. However, the Rapidlab 865 and Rapidlab 405 analyzers measured significantly reduced amounts of ionized calcium after specimens were spiked with perchlorate, and the ABL 815 flex measured significantly reduced amounts of ionized calcium when the spiked perchlorate concentrations were above 0.5 mmol/L. When specimens were spiked with additional ionized calcium, but not perchlorate, both the Rapidlab 865 and ABL 815 flex were able to detect hypercalcaemia, but with a 1.4 mmol/L spiked perchlorate concentration, Rapidlab 865 failed to detect hypercalcaemia and ionized calcium measurements measured by ABL 815 flex decreased. The authors state that spiked perchlorate was stable in anticoagulated blood specimens at room temperature for 48 h and that laboratory prepared stock solutions of sodium perchlorate and Irenat sodium perchlorate had identical effects on ionized calcium readings using the Rapidlab 865 and the ABL 815 flex.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerSmall moleculeIon selective electrodePre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Biospecimen components Multiple concentrations evaluatedSpiked calcium chlorideSpiked sodium perchlorate (Irenat)Spiked sodium perchlorate (laboratory prepared)Storage Time at room temperature 0 h24 h48 hClinical chemistry/auto analyzer Specific Technology platform ABL 815 flexCobas b221GEM 4000Rapidlab 405Rapidlab 865