Perchlorate (Irenat®) may falsely lower measured ionised calcium.

Gruber M, Nehring C, Creutzenberg M, Graf B, Hopf S

Publication: Clin Chem Lab Med (2011) • Vol 49 • Page 1019-24PMID: 21428857View on PubMed

Purpose

The purpose of this paper was to compare native and spiked ionized calcium measurements between 5 blood gas analyzers after blood specimens were spiked with perchlorate.

Conclusion

The measurement of ionized calcium was falsely decreased in specimens containing perchlorate when measured by the Rapidlab 865 and 405 blood gas analyzers, as well as the ABL 815 flex, but not the GEM 4000 and cobas b221 blood gas analyzers.

Studies (1)

Purpose

The purpose of this study was to compare ionized calcium measurements between 5 blood gas analyzers in blood spiked with calcium chloride and/or different concentrations of perchlorate. In addition, the stability of perchlorate in blood stored at room temperature was examined.

Summary of Findings

The measurement of native ionized calcium concentration was unaffected by the addition of perchlorate to blood specimens using the GEM 4000 and cobas b221 blood gas analyzers. However, the Rapidlab 865 and Rapidlab 405 analyzers measured significantly reduced amounts of ionized calcium after specimens were spiked with perchlorate, and the ABL 815 flex measured significantly reduced amounts of ionized calcium when the spiked perchlorate concentrations were above 0.5 mmol/L. When specimens were spiked with additional ionized calcium, but not perchlorate, both the Rapidlab 865 and ABL 815 flex were able to detect hypercalcaemia, but with a 1.4 mmol/L spiked perchlorate concentration, Rapidlab 865 failed to detect hypercalcaemia and ionized calcium measurements measured by ABL 815 flex decreased. The authors state that spiked perchlorate was stable in anticoagulated blood specimens at room temperature for 48 h and that laboratory prepared stock solutions of sodium perchlorate and Irenat sodium perchlorate had identical effects on ionized calcium readings using the Rapidlab 865 and the ABL 815 flex.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeIon selective electrode
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen components
    Multiple concentrations evaluated
    Spiked calcium chloride
    Spiked sodium perchlorate (Irenat)
    Spiked sodium perchlorate (laboratory prepared)
    StorageTime at room temperature
    0 h
    24 h
    48 h
    Clinical chemistry/auto analyzer SpecificTechnology platform
    ABL 815 flex
    Cobas b221
    GEM 4000
    Rapidlab 405
    Rapidlab 865