The effect of preparation, storage and shipping of dried blood spots on the activity of five lysosomal enzymes.
Elbin CS, Olivova P, Marashio CA, Cooper SK, Cullen E, Keutzer JM, Zhang XK
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine the effects of DBS specimen preparation methods on the measurement of GAA, GLA, GBA, acid sphingomyelinase (ASM), and galactocerebrosidase (GALC) by MS/MS and fluorometric analysis (GAA only).
Summary of Findings
For one of 3 donors, use of lithium heparin as opposed to EDTA as an anticoagulant resulted in much higher within run CVs and assay imprecision for MS/MS analysis (14.3%-58.4% compared to 5.4%-14.5% for EDTA-blood). No consistent differences were observed between EDTA and lithium heparin-blood for the other two donors. MS/MS measured enzyme activity tended to increase as the volume of blood spotted increased, but changes were minor. Preparing DBS by sampling from the top of unmixed blood that sat undisturbed for 2 hours resulted in significantly elevated activities for all 5 enzymes measured by MS/MS. On the other hand, sampling from the bottom of unmixed blood that sat undisturbed for 4 hours resulted in significantly decreased enzyme activity for all 5 enzymes. The effects of blood volume and settling were comparable for GAA activity measured by MS/MS and the enzyme assay. Enzyme activity was significantly increased when DBS were prepared on FTA Classic cards compared to 903 Protein Saver cards for GAA (p<0.001), GBA (p<0.001), and GLA (p<0.04), but not for GALC or ASM where activity was insignificantly lower. Enzyme activity was not significantly different between DBS on 903 Protein Saver cards and Ahlstrom 226 paper. Enzyme activity was significantly decreased for all 5 enzymes when blood spots were dried at 37 or 45 degrees C and 74-83% humidity compared to drying at 4 or 18-20 degrees C with lower humidity. For GAA, similar effects of increased drying temperature and humidity were observed for MS/MS analysis and the enzyme assay.
Biospecimens
FluidBloodPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
ProteinMS/MSProteinEnzyme assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTALithium heparinBiospecimen Acquisition Type of collection container/solution Ahlstrom 226 paperFTA cardProtein Saver cardBiospecimen Aliquots and Components Aliquot size/volume 100 uL125 uL25 uL50 uL75 uLBiospecimen Aliquots and Components Biospecimen mixing Inverted 5-10 timesSampled from bottom of tubeSampled from top of tubeUnmixed and undisturbed for 2 hUnmixed and undisturbed for 4 hBiospecimen Preservation Conditions during fixation/preservation 21-26% humidity30% humidity74-83% humidityBiospecimen Preservation Temperature of fixation/preservation 18-20 degrees C37 degrees C4 degrees C45 degrees CMS/MS Specific Technology platform Enzyme assay
Purpose
The purpose of this study was to determine the effects of DBS storage and shipping on the measurement of GAA, GLA, GBA, ASM, and by MS/MS and fluorometric analysis (GAA only).
Summary of Findings
Storage of DBS at 37 degrees C, as opposed to -20 degrees C, was generally detrimental to enzyme activity, and effects increased with storage duration. However, storage of DBS in plastic bags with desiccant attenuated these effects compared to storage in paper envelopes. Similar effects of DBS storage temperature and container were observed for GAA activity with MS/MS analysis and the enzyme assay. Shipping DBS at ambient summer temperatures for 9 days generally resulted in minor increases in GAA, GLA, and GBA activity and minor decreases in ASM and GALC activity, regardless of storage container.
Biospecimens
FluidBloodPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
ProteinMS/MSProteinEnzyme assayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Between site transportation method MailedNot transportedStorage Storage duration 1 d2 d3 d9 dStorage Storage temperature -20 degrees C37 degrees CStorage Type of storage container Paper envelopePlastic bag with desiccantPlastic bag without desiccantMS/MS Specific Technology platform Enzyme assay