The effect of preparation, storage and shipping of dried blood spots on the activity of five lysosomal enzymes.

Elbin CS, Olivova P, Marashio CA, Cooper SK, Cullen E, Keutzer JM, Zhang XK

Publication: Clin Chim Acta (2011) • Vol 412 • Page 1207-12PMID: 21419758View on PubMed

Purpose

The purpose of this paper was to determine the effects of dried blood spot (DBS) specimen preparation method, storage, and shipping on the measurement of lysosomal enzymes by tandem mass spectrometry (MS/MS) and fluorometric analysis.

Conclusion

Anticoagulation with lithium heparin resulted in higher assay imprecision in one specimen compared to EDTA for MS/MS analysis. Enzyme activity tended to increase as the volume of blood spotted increased, but changes were minor. Sampling from the top of unmixed blood resulted in increased enzyme activity while sampling from the bottom resulted in decreased enzyme activity. DBS on FTA Classic cards had increased acid alpha-glucosidase (GAA), acid beta-glucocerebrosidase (GBA), and acid alpha-galactosidase (GLA) activity compared to 903 Protein Saver cards, while no significant differences were noted between Protein Saver cards and Ahlstrom 226 paper. Higher temperature and humidity during DBS drying decreased enzyme activity. Storage of DBS at 37 degrees C, as opposed to -20 degrees C, was generally detrimental to enzyme activity, and effects increased with storage duration. However, storage of DBS in plastic bags with desiccant attenuated these effects compared to storage in paper envelopes. Shipping DBS at ambient summer temperatures for 9 days resulted in only minor changes in enzyme activity. The effects of blood volume, settling, drying temperature and humidity, and storage on GAA activity were comparable when measured by MS/MS and the enzyme assay.

Studies (2)

Purpose

The purpose of this study was to determine the effects of DBS specimen preparation methods on the measurement of GAA, GLA, GBA, acid sphingomyelinase (ASM), and galactocerebrosidase (GALC) by MS/MS and fluorometric analysis (GAA only).

Summary of Findings

For one of 3 donors, use of lithium heparin as opposed to EDTA as an anticoagulant resulted in much higher within run CVs and assay imprecision for MS/MS analysis (14.3%-58.4% compared to 5.4%-14.5% for EDTA-blood). No consistent differences were observed between EDTA and lithium heparin-blood for the other two donors. MS/MS measured enzyme activity tended to increase as the volume of blood spotted increased, but changes were minor. Preparing DBS by sampling from the top of unmixed blood that sat undisturbed for 2 hours resulted in significantly elevated activities for all 5 enzymes measured by MS/MS. On the other hand, sampling from the bottom of unmixed blood that sat undisturbed for 4 hours resulted in significantly decreased enzyme activity for all 5 enzymes. The effects of blood volume and settling were comparable for GAA activity measured by MS/MS and the enzyme assay. Enzyme activity was significantly increased when DBS were prepared on FTA Classic cards compared to 903 Protein Saver cards for GAA (p<0.001), GBA (p<0.001), and GLA (p<0.04), but not for GALC or ASM where activity was insignificantly lower. Enzyme activity was not significantly different between DBS on 903 Protein Saver cards and Ahlstrom 226 paper. Enzyme activity was significantly decreased for all 5 enzymes when blood spots were dried at 37 or 45 degrees C and 74-83% humidity compared to drying at 4 or 18-20 degrees C with lower humidity. For GAA, similar effects of increased drying temperature and humidity were observed for MS/MS analysis and the enzyme assay.

  • Biospecimens

    FluidBlood
  • Preservation Types

    Other Preservative
  • Diagnoses

    Not specified
  • Platforms

    ProteinMS/MS
    ProteinEnzyme assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    EDTA
    Lithium heparin
    Biospecimen AcquisitionType of collection container/solution
    Ahlstrom 226 paper
    FTA card
    Protein Saver card
    Biospecimen Aliquots and ComponentsAliquot size/volume
    100 uL
    125 uL
    25 uL
    50 uL
    75 uL
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    Inverted 5-10 times
    Sampled from bottom of tube
    Sampled from top of tube
    Unmixed and undisturbed for 2 h
    Unmixed and undisturbed for 4 h
    Biospecimen PreservationConditions during fixation/preservation
    21-26% humidity
    30% humidity
    74-83% humidity
    Biospecimen PreservationTemperature of fixation/preservation
    18-20 degrees C
    37 degrees C
    4 degrees C
    45 degrees C
    MS/MS SpecificTechnology platform
    Enzyme assay