Pseudohypocalcemia caused by perchlorate (Irenat).
Durner J, Winkler-Budenhofer U, Gahr S, Samtleben W, Schönermarck U
Purpose
The purpose of this paper was to determine the effects of perchlorate on the measurement of ionized calcium by the RAPIDpoint 405 blood gas analyzer (BGA).
Conclusion
Significantly lower amounts of ionized calcium were measured with BGA compared to the standard laboratory method (Roche AVL 9180 Electrolyte Analyzer) for both whole blood spiked with perchlorate and blood from patients receiving perchlorate treatment.
Studies (1)
Purpose
The purpose of this study was to determine the effects of perchlorate on the measurement of ionized calcium by BGA using spiked blood specimens and specimens from patients with renal disease receiving perchlorate.
Summary of Findings
Significantly lower amounts of ionized calcium were measured with BGA compared to the standard method (Roche AVL 9180 Electrolyte Analyzer) after whole blood specimens from 2 donors were spiked with perchlorate. In addition, for 3 patients receiving perchlorate treatment, ionized calcium measured by BGA significantly decreased during treatment compared to measurements taken before and after treatment. Ionized calcium measurements taken with the Roche AVL did not decrease during treatment compared to before or after treatment.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalOther diagnosesPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 1.20 mmol/L ionized calcium1.26 mmol/L ionized calciumPreaquisition Diagnosis/ patient condition Renal diseasePreaquisition Other drugs NonePerchlorateBiospecimen Acquisition Time of biospecimen collection After perchlorate treatmentBefore perchlorate treatmentDuring perchlorate treatmentBiospecimen Aliquots and Components Biospecimen components 0.143 mmol/L spiked perchlorate0.29 mmol/L spiked perchlorate0.6 mmol/L spiked perchlorate1.1 mmol/L spiked perchlorate1.7 mmol/L spiked perchlorate2.3 mmol/L spiked perchlorateClinical chemistry/auto analyzer Specific Technology platform RAPIDpoint 405 Blood Gas AnalyzerRoche AVL 9180 Electrolyte Analyzer