Skip to main content

Use of the BD vacutainer rapid serum tube reduces false-positive results for selected beckman coulter Unicel DxI immunoassays.

Strathmann FG, Ka MM, Rainey PM, Baird GS

Publication: Am J Clin Pathol (2011) • Vol 136 • Page 325-9PMID: 21757607View on PubMed

Purpose

The purpose of this paper was to determine the effects of collecting blood in rapid serum tubes (RST) rather than lithium-heparin plasma tubes (PST) on the measurement of clinical chemistry analytes.

Conclusion

Most analytes were unaffected by collection tube type, but values for thyroid-stimulating hormone (TSH), human chorionic gonadotropin (hCG), creatine kinase-MB isoenzyme (CK-MB), and cardiac troponin I (TnI) were often elevated in specimens collected in PST.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collecting blood in RST rather than PST, and repeat centrifugation on the measurement of 27 clinical chemistry analytes.

Summary of Findings

When specimens collected in RST and PST were compared, values for sodium, potassium, chloride, total carbon dioxide, anion gap, total calcium, glucose, blood urea nitrogen, creatinine, total protein, albumin, phosphorus, magnesium, total bilirubin, direct bilirubin, aspartate aminotransferase, alanine aminotransferase, alkaline phosphatase, amylase, pancreatic amylase, lipase, ethanol, and acetaminophen correlated well (r2>0.95), but clinically significant differences in TnI, CK-MB, hCG and TSH were sometimes seen during the initial analysis. Recentrifugation generally did not affect results for discrepant specimens collected in RST, but in one specimen with an initially high hCG level, repeat centrifugation resulted in a normal result. In contrast, after recentrifugation of specimens collected in PST, lower concentrations of TnI, CK-MB, hCG and TSH were often seen which were more similar to those of specimens collected in RST. In conclusion, collection of blood into RST decreases the number of false positive results for TnI, CK-MB, hCG and TSH compared to collection in PST.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Cardiovascular Disease
    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Lithium heparin plasma separator tube
    Vacutainer rapid serum tube
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Different number of centrifugation steps compared