Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus.
Waring WS, Evans LE, Kirkpatrick CT
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of NaF-KOx and storage on the measurement of glucose in blood specimens.
Summary of Findings
Glucose concentrations were 4.7% lower when blood was collected in NaF-KOx tubes and stored 6.5-9.5 h rather than collected in serum gel tubes and separated within 15 min (p<0.0001). The difference in glucose concentrations between tubes was not affected by storage duration at room temperature (p=0.100). The authors believe the decrease in blood glucose concentrations found when blood was collected in NaF-KOx could alter the diabetes diagnosis rate.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Other PreservativeDiagnoses
NormalPlatforms
CarbohydrateClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution NaF-KOx tubeSerum gel tubeBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedBiospecimen Preservation Type of fixation/preservation None (fresh)Potassium oxalateSodium fluorideStorage Storage duration 15 min6.5-9.5 h