Errors due to heparin in the estimation of plasma sodium and potassium concentrations.
Shek CC, Swaminathan R
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to compare the effects of anticoagulation with sodium heparin rather than lithium heparin on the measurement of sodium and potassium in plasma by flame photometry and in whole blood by ion selective electrode.
Summary of Findings
When blood was collected into tubes containing 100 ul, but not 20 ul of 5000 U/mL sodium heparin, sodium and potassium, measured by ion selective electrode in plasma, were decreased (both p<0.001), and sodium, measured by flame photometry in whole blood, was increased (p<0.001) compared to levels in specimens anticoagulated with lithium heparin. Collection of blood into tubes containing 100 ul of 1000 U/mL sodium heparin increased sodium and decreased potassium levels in plasma and in whole blood compared to lithium heparin specimens (all p<0.001).
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Electrolyte/MetalFlame emission photometryElectrolyte/MetalIon selective electrodePre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Lithium heparinMultiple concentrations evaluatedSodium heparinBiospecimen Aliquots and Components Blood and blood products PlasmaWhole blood