Serum collected with lithium iodoacetate is unsuitable for electrolyte and lactate dehydrogenase measurements.

Robertson EA, Chesler RA, Elin RJ

Publication: Clin Chem (1980) • Vol 26 • Page 1228-9PMID: 7389098View on PubMed

Purpose

The purpose of this paper was to determine the effects of lithium iodoacetate, specimen volume and analysis method on measurement of electrolytes and lactate dehydrogenase (LD) in blood specimens.

Conclusion

When blood was collected in tubes containing lithium iodoacetate, more chloride, potassium, and LD, a higher rate of hemolysis, and less sodium were measured when compared to blood collected in serum separator tubes (SST), but collection tube type had no effect on measured carbon dioxide levels. The effect of lithium iodoacetate decreased when 10 mL blood specimens were collected instead of 3 mL. The increase in measured chloride when lithium iodoacetate tubes were used was larger when measured using the PVA-4 analyzer instead of SMAC.

Studies (1)

Purpose

The purpose of this study was to determine the effects of anticoagulant, specimen volume and analysis method on measurement of electrolytes and LD in blood specimens.

Summary of Findings

When 3 mL of blood was collected in tubes containing lithium iodoacetate, more chloride (p<0.0001), potassium (p<0.001), and LD (p<0.0001) and less sodium (p<0.0001) were measured when compared to blood collected in SST. These effects were attenuated when a 10 mL draw volume was used instead of 3 mL but were still significant for all analytes except, carbon dioxide and sodium measured using the PVA-4 photovolt analyzer. There were no differences in measured carbon dioxide levels between lithium iodoacetate tubes and SST. The authors believe the increases in potassium and LD were most likely due to approximately 100 fold more hemolysis in specimens collected in tubes containing lithium iodoacetate (p<0.01). The increase in chloride with lithium iodoacetate tubes was larger when measured using the PVA-4 analyzer compared to SMAC. Furthermore, adding sodium iodide to sera increased measured chloride using the PVA-4 photovolt analyzer. The authors conclude that blood specimens collected in tubes containing lithium iodoacetate are not suitable for analysis of electrolytes and LD, particularly using the PVA-4 photovolt analyzer.

  • Biospecimens

    FluidPlasma
    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinClinical chemistry/auto analyzer
    ProteinHematology/ auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    MorphologyMacroscopic observation
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    10 mL
    3 mL
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Clinical chemistry/auto analyzer SpecificTechnology platform
    PVA-4 Photovolt analyzer
    Technicon SMAC