Influence of Storage Temperature and High-Temperature Antigen Retrieval Buffers

Grabau D, Nielsen O, Hansen S, Nielsen MM, Laenkholm A, Knoop A, Pfeiffer P

Publication: Applied Immunohistochemistry (1998) • Vol 6 • Page 209-13

Purpose

The purpose of this paper was to determine the stability of clinically relevant antigens following prolonged storage of slide-mounted formaldehyde-fixed, paraffin-embedded (FFPE) tissue sections at different temperatures.

Conclusion

Storage temperature adversely affected the antigenicity of stored slide-mounted FFPE tissue sections for all antigens investigated regardless of antigen retrieval method, with the exception of c-erbB2 which remained stable in all sections examined. While the extent of immuno-positive staining was similar among tissue sections stored for 3 years at -80 degrees C and those that were freshly cut, sections stored at 20 degrees C for 3 years exhibited a 50% decline in immuno-positive staining compared to freshly sectioned controls for ER, Ki67, RB1, E-cadherin, and EGFR, and a slightly greater loss for p53 and bcl2. Antigen retrieval in TEG buffer for 25 or 35 min, compared to citrate buffer for 25 min, increased the extent of immunostaining among both control and stored tissue sections; however, in relation to identically processed controls improvements among stored sections were modest, and storage temperature- and duration- dependent effects were still present.

Studies (1)

Purpose

The purpose of this study was to determine the stability of clinically relevant antigens in FFPE slide-mounted tissue sections that were stored at -80, 4, or 20 degrees C for 3 years or 1 day prior to analysis. The influence of different antigen retrieval solutions and durations on immunostaining was also investigated. Two lung and four breast carcinoma specimens that were fixed in 4% neutral buffered formaldehyde for 24-48 h were analyzed.

Summary of Findings

The authors note that statistical significance was not assessed due to small sample size. Storage temperature adversely affected the antigenicity of stored slide-mounted formaldehyde-fixed, paraffin-embedded tissue sections for all antigens investigated regardless of antigen retrieval method, with the exception of c-erbB2 which remained stable in all sections examined. While the extent of immuno-positive staining was similar among tissue sections stored for 3 years at -80 degrees C and those that were freshly cut, sections stored at 20 degrees C for 3 years exhibited a 50% decline in immuno-positive staining compared to freshly sectioned controls for ER, Ki67, RB1, E-cadherin, and EGFR, and a slightly greater loss for p53 and bcl2. Antigen retrieval in TEG buffer for 25 or 35 min, compared to citrate buffer for 25 min, increased the extent of immunostaining among both control and stored tissue sections; however, in relation to identically processed controls improvements among stored sections were modest, and storage temperature- and duration- dependent effects were still present.

  • Biospecimens

    TissueBreast
    TissueLung
  • Preservation Types

    Other Preservative
  • Diagnoses

    NeoplasticCarcinoma
  • Platforms

    ProteinImmunohistochemistry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    1 d
    3 y
    StorageStorage temperature
    -80 degrees C
    20 degrees C
    4 degrees C
    Analyte Extraction and PurificationAntigen retrieval
    Citrate buffer, pH 6 for 25 min
    TEG buffer, pH 9 for 25 min
    TEG buffer, pH 9 for 35 min
    Immunohistochemistry SpecificTargeted peptide/protein
    E-cadherin
    EGFR
    ER
    Ki67
    RB1
    bcl2
    c-erbB2
    p53