Evaluation of the i-STAT Portable Clinical Analyzer for point-of-care blood testing in the intensive care units of a university children's hospital.

Papadea C, Foster J, Grant S, Ballard SA, Cate JC 4th, Southgate WM, Purohit DM

Publication: Ann Clin Lab Sci (2002) • Vol 32 • Page 231-43PMID: 12175085View on PubMed

Purpose

The purpose of this paper was to investigate the reliability of the iSTAT portable clinical analyzer for analysis of pediatric blood specimens.

Conclusion

The iSTAT was found to compare well with other analyzers for analysis of all analytes except creatinine.

Studies (1)

Purpose

The purpose of this study was to determine the effects of analyzer type on clinical chemistry and hematology analysis of pediatric blood specimens.

Summary of Findings

The day to day coefficient of variance (CV) using iSTAT for electrolytes, glucose, urea nitrogen and pH was less than 2% and the CV for oxygen and carbon dioxide were less than 6.5%, but the hematocrit and creatinine were highly variable with CVs of 21 and 23.6%, respectively. The CVs using iSTAT were comparable to the CVs using the VITROS or ABL analyzers for all analytes except creatinine and hematocrit which had CVs of <3% using the VITROS analyzer and <9.1% using the ABL analyzer, respectively. The different cartridges performed similarly for all of the analytes that were analyzed by multiple cartridge types. However, a follow-up study of creatinine levels found less correlation between VITROS values using a new lot of cartridges, resulting in possible clinical differences. In specimens analyzed using iSTAT, a small proportional negative bias was found for ionized calcium (iCa) and partial pressure oxygen (PO2), and a there was a positive bias for creatinine. The iSTAT analyzer was found to have good reproducibility for children under 3 months old and those over 3 months old. The authors conclude that iSTAT is suitable for clinical chemistry and hematology analysis for all analytes except creatinine.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    Small moleculepH
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionPatient age
    < 3 months
    > 3 months
    Clinical chemistry/auto analyzer SpecificDetection method
    Creatinine cartridge
    EC6+ cartridge
    EC8+ cartridge
    G3+ cartridge
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Coulter STKS hematology analyzer
    Radiometer ABL 725 blood gas analyzer
    Vitros 750 chemistry analyzer
    iSTAT portable clinical analyzer
    Hematology/ auto analyzer SpecificTechnology platform
    Coulter STKS hematology analyzer
    iSTAT portable clinical analyzer