Heparinized "microtainer" tube evaluated for collection of capillary blood.

Norling LL, Smith CH, Landt M

Publication: Clin Chem (1984) • Vol 30 • Page 810-1PMID: 6713651View on PubMed

Purpose

The purpose of this paper was to determine the effects of collection tube type on hemolysis, plasma yield and 8 clinical chemistry analytes.

Conclusion

There were no clinically relevant effects of collecting blood in caraway rather than microtainer tubes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type on hemolysis, plasma yield and 8 clinical chemistry analytes.

Summary of Findings

The mean sodium concentration was 138 mmol/L when specimens were collected in Caraway tubes versus 136 mmol/L when collected in microtainer tubes (p<0.05), but this difference was not clinically relevant. A significant loss of CO2 was observed in specimens collected in microtainer tubes with storage prior to analysis of longer than 4 h (p<0.001). A non-significant reduction in hemolysis was observed in specimens collected in microtainer rather than Caraway tubes. Collection tube type had no effect on plasma yield, potassium, carbon dioxide, chloride, glucose, urea, total bilirubin, or creatinine.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalFlame emission photometry
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Small moleculeSpectrophotometry
    CarbohydrateClinical chemistry/auto analyzer
    MorphologyHematology/ auto analyzer
    MorphologyMacroscopic observation
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Caraway tube
    Microtainer tube
    StorageStorage duration
    Less than 4 h
    Less than 6 h