Interference in ionized calcium measurements by heparin salts.
Landt M, Hortin GL, Smith CH, McClellan A, Scott MG
Purpose
The purpose of this paper was to determine the effects of anticoagulant type and concentration on measured calcium, pH and gas levels in blood specimens.
Conclusion
Lithium (Li) heparin and a high concentration of Zn (Zn) heparin decreased ionized calcium (iCa), while a low concentration of Zn heparin increased iCa compared to the control of freshly drawn whole blood. When 0.75-3 mL of blood was drawn into tubes containing calcium-neutralized Zn/lithium (CNLZ) heparin with a Zn concentration of 74.2 g/kg iCa, pH, oxygen, carbon dioxide and total calcium levels were similar to those seen in the control. Collection in tubes containing electrolyte balanced (EB) heparin led to increased measured total calcium, but iCa, pH, oxygen and carbon dioxide were unaffected. Although iCa differed between serum and CNLZ or EB plasma collected from cardiac patients, total calcium levels were similar in CNLZ plasma and serum, but were significantly lower (p<0.01) than levels measured in EB plasma.
Studies (1)
Purpose
The purpose of this study was to determine the effects of anticoagulant type and concentration on iCa, pH and gas levels in blood specimens. The effect of anticoagulant concentration was examined by partial filling of syringes and by using syringes with less anticoagulant.
Summary of Findings
At all concentrations, Li heparin decreased iCa below levels found in control blood without anticoagulant. At low concentrations, Zn heparin increased iCa levels, but at higher concentrations, Zn heparin led to decreased iCa and pH. In contrast, EB heparin had no effect on iCa measurement. When 1.5 mL or less blood was collected in tubes containing CNLZ heparin with more than 78 g/ kg Zn, iCa concentrations were higher than in control blood. When CNLZ heparin tubes contained less than 59 g/kg Zn, iCa measurements decreased in specimens with a volume between 0.75 and 3 mL. However, iCa values were within 0.025 mmol/L of control when 0.75-3 mL specimens were collected in CNLZ heparin tubes with a Zn concentration between 59.4 and 78 g/kg. Collection of blood into tubes containing CNLZ heparin with a Zn concentration of 74.2 g/kg led to pH, oxygen, carbon dioxide and total calcium levels similar to control, but collection in tubes containing EB heparin led to an increase in measured total calcium, but no other significant effects. Although iCa differed between serum and CNLZ or EB plasma collected from cardiac patients, total calcium levels were similar in CNLZ plasma and serum, but were significantly lower (p<0.01) than in EB plasma.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerSmall moleculepHPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Calcium-neutralized zinc/lithium heparinElectrolyte balanced heparinLithium heparinMultiple concentrations evaluatedNoneZinc heparinBiospecimen Aliquots and Components Aliquot size/volume 0.4-3 mL