Skip to main content

A comparison of routine and rapid microwave tissue processing in a surgical pathology laboratory. Quality of histologic sections and advantages of microwave processing.

Rohr LR, Layfield LJ, Wallin D, Hardy D

Publication: Am J Clin Pathol (2001) • Vol 115 • Page 703-8PMID: 11345834View on PubMed

Purpose

The purpose of this paper was to determine the effects of microwave versus conventional tissue processing for small and large specimens from a variety of tissues.

Conclusion

There were no significant differences between microwave processed and conventionally processed specimens. The microwave processor significantly decreased the amount of time required for processing and eliminated the need for a clearing step with xylene.

Studies (1)

Purpose

The purpose of this study was to determine the effects of microwave versus conventional tissue processing schedules for small and large specimens from several different tissues on morphology. Small specimens were processed using shortened microwave and vacuum tissue processing schedules.

Summary of Findings

There were no significant differences in morphology between microwave processed and conventionally processed specimens. The microwave processor significantly decreased the amount of time required for processing and eliminated the need for a clearing step with xylene by boiling out residual isopropanol through the use of a higher paraffin impregnation temperature. Of the 158 paired specimens processed by both a microwave processor and a vacuum tissue processor, 6 microwave processed specimens were deemed suboptimal but satisfactory for diagnosis and 1 was deemed unsatisfactory for diagnosis by two independent, blinded pathologists. The majority of these specimens showed inadequate fixation. On the other hand, 8 specimens processed in the vacuum tissue processor were deemed suboptimal but satisfactory for diagnosis. The problems encountered with these specimens included lack of cellular detail and sections that were too thick.

  • Biospecimens

    TissueColorectal
    TissueGall Bladder
    TissueUterus
    TissueSmall Bowel
    TissueBreast
    TissueProstate
    TissueLymph Node
    TissueLiver
    TissueEndometrium
    TissueCervix
  • Preservation Types

    Formalin
  • Diagnoses

    Not specified
  • Platforms

    MorphologyH-and-E microscopy
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    Large specimen (>2 mM thick or with abundant blood/mucus/fat)
    Small specimen (<2 mM thick and <10 mM diameter)
    Biospecimen PreservationClearing agent
    None
    Xylene
    Biospecimen PreservationDehydration reagent
    Absolute alcohol
    Graded alcohols
    Isopropanol
    Biospecimen PreservationDuration of tissue/ specimen processing
    12 h (vacuum processor)
    15 min (microwave)
    60 min (microwave)
    8 h (vacuum processor)
    90 min (microwave)
    Biospecimen PreservationEmbedding duration/condition
    60 degrees C
    75-80 degrees C