Evaluation and validation of total RNA extraction methods for microRNA expression analyses in formalin-fixed, paraffin-embedded tissues.

Doleshal M, Magotra AA, Choudhury B, Cannon BD, Labourier E, Szafranska AE

Publication: J Mol Diagn (2008) • Vol 10 • Page 203-11PMID: 18403610View on PubMed

Purpose

The purpose of this paper was to determine the effects of RNA extraction method on total RNA and microRNA yield and amplification success.

Conclusion

The RecoverAll kit was the best in terms of RNA yield, purity, consistency and real-time PCR success from formalin-fixed paraffin-embedded (FFPE) tissue blocks, but an average or 1.65 fold less miRNA was detected in FFPE specimens extracted using the RecoverAll kit than in frozen specimens eetracted with the mirVana kit. RNA yield increased linearly with section number for up to 12 sections, as long as additional proteinase K was added to sections with a cross-sectional area of >150 mm2. The RNA yield and cycle threshold (CT) using the RecoverAll kit was highly reproducible within a tissue and between operators.

Studies (1)

Purpose

The purpose of this study was to determine the effects of RNA extraction method and block size on total RNA and microRNA yield and amplification success from FFPE blocks of 10 different tissues. The FFPE blocks used in this study were stored for up to 12 years.

Summary of Findings

When RNA was extracted from lung and breast cancers using the RecoverAll and RNeasy kit there was a much higher RNA (2-3 fold) and microRNA (5-20 fold) yield than when the other kits were used. When RNA was isolated from 13 additional FFPE blocks that had been stored for up to 12 y, the RecoverAll kit produced a higher, purer and more consistent yields than the RNeasy kit, but there was no difference in the electropherograms. miR-24 and miR-103 were detected at slightly higher levels (2 fold) and miR-191 at comparable levels when RNA was extracted using the RecoverAll kit rather than the RNeasy kit. The 18S ribosomal RNA and hGUSB were also detected at higher levels (11.7 and 2.4 fold, respectively) when the RecoverAll kit was used, but due to variability between the tissues, the effect was not significant. On average, 1.65 fold less miRNA was detected in FFPE specimens extracted using the RecoverAll kit than in frozen specimens when extraction was performed with the mirVana kit. RNA yield increased linearly with section number for up to 12 sections, as long as additional proteinase K was added to sections with a cross-sectional area of >150 mm2. Finally, the RNA yield and CT using the RecoverAll kit was highly reproducible within a tissue and between operators.

  • Biospecimens

    TissueCervix
    TissueBreast
    TissueGall Bladder
    TissueLung
    TissueKidney
    TissueTestis
    TissueStomach
    TissueUterus
    TissueProstate
    TissueSpleen
    TissueLiver
  • Preservation Types

    FormalinFrozen
  • Diagnoses

    Neoplastic
    NeoplasticNormal Adjacent
    NeoplasticCarcinoma
    Not specified
  • Platforms

    RNAReal-time qRT-PCR
    RNASpectrophotometry
    RNAAutomated electrophoresis/Bioanalyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    >150 mm2
    12 sections
    16 sections
    4 sections
    8 sections
    <75 mm2
    Biospecimen PreservationType of fixation/preservation
    Formalin (buffered)
    Frozen
    StorageStorage duration
    1-12 years
    Analyte Extraction and PurificationAnalyte isolation method
    Absolutely RNA FFPE kit
    High Pure FFPE RNA micro kit
    PureLink FFPE RNA isolation kit
    RecoverAll Total Nucleic Acid isolation kit
    Rneasy FFPE kit
    mirVana miRNA isolation kit
    Real-time qRT-PCR SpecificTargeted nucleic acid
    18S rRNA
    hGUSB
    miR-103
    miR-191
    miR-24