Stability of serum samples and hemolysis interference on the high sensitivity troponin T assay.
Li A, Brattsand G
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of hemolysis and storage of blood specimens at room temperature on serum troponin T levels as measured by the hs assay. Hemolysis free serum pools with troponin T were added to blood for non-hemolysed specimens, and hemolysate was added to serum specimens containing troponin T to resemble hemolysed specimens.
Summary of Findings
Storage of blood specimens at room temperature had no significant effect on serum troponin T levels in the absence of hemolysis. However, in the presence of hemolysis, storage at room temperature resulted in negative interference with the troponin T hs assay in a hemoglobin concentration-dependent and time-dependent, but troponin T concentration-independent manner. However, storage of hemolysed specimens at room temperature for up to 6 hours resulted in changes in troponin T measurements of less than 10%, except at the highest degree of hemolysis (2 g/L of added hemoglobin).
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution SST II Advance tubesBiospecimen Aliquots and Components Biospecimen components 0 g/L hemoglobin0-0.12 g/L hemoglobin0.5 g/L hemoglobin1 g/L hemoglobin2 g/L hemoglobin35 ng/L troponin T43 ng/L troponin T480 ng/L troponin T80 ng/L troponin TBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedBiospecimen Aliquots and Components Hemolysis Freeze/thaw-inducedHemolysate addedNo hemolysate addedStorage Time at room temperature 0 h15 min24 h4 h6 h8 h