Suitability of collection tubes with separator gels for therapeutic drug monitoring.
Koch TR, Platoff G
Purpose
The purpose of this paper was to determine the effects of storage in different collection tubes containing separation gel on the levels of therapeutic drugs in serum.
Conclusion
When serum was transferred to Corvac tubes, no differences were found in the initial levels of any of the therapeutic drugs, but lidocaine levels decreased significantly by 24 h. Serum transferred to Autosep tubes initially had falsely increased theophylline concentrations. When Autosep tubes were filled half-way with serum initial amikacin levels were elevated and pentobarbital levels depressed. After storage for 72 h in a half-full Autosep tube, digoxin levels were elevated, and lidocaine levels depressed. Transfer of serum to BD serum separator tube (SST) initially increased digoxin and decreased phenytoin levels, but with storage phenytoin, lidocaine and quinidine levels decreased. The authors conclude that Autosep and Corvac tubes are suitable for therapeutic drug monitoring, but tubes must be filled completely and analyzed within a few hours.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storage in four different collection tubes containing separation gel on the levels of 15 therapeutic drugs in serum. The effect of partially filling Autosep tubes was also examined. Spiked serum was added to each tube.
Summary of Findings
When serum was transferred to Corvac tubes, no differences were found in the initial levels of any of the therapeutic drugs, but lidocaine levels decreased significantly by 24 h. Serum transferred to Autosep tubes initially had falsely increased theophylline concentrations. When Autosep tubes were filled half-way with serum initial amikacin levels were elevated and pentobarbital levels depressed. After storage for 72 h in a half-full Autosep tube, digoxin levels were elevated, and lidocaine levels depressed. Transfer of serum to SST initially increased digoxin and decreased phenytoin levels, but with storage phenytoin, lidocaine and quinidine levels decreased. Levels of phenobarbital, primidone, carbamazepine, valproic acid, methotrexate, gentamicin, tobramycin, and chloramphenicol were not affected by tube type or storage. The authors conclude that Autosep and Corvac tubes are suitable for therapeutic drug monitoring, but tubes must be filled completely and analyzed within a few hours.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerSmall moleculeImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Aliquot size/volume FullHalf-fullStorage Storage duration 0 h24 h6 h72 hStorage Storage temperature 4 degrees CStorage Type of storage container BD SSTMonoject Corvac tubeTerumo Autosep tubeTerumo Venoject tube