Skip to main content

Evaluation of a new blood-collecting device ("microtainer") that is suited for pediatric use.

Hicks JM, Rowland GL, Buffone GJ

Publication: Clin Chem (1976) • Vol 22 • Page 2034-6PMID: 1000802View on PubMed

Purpose

The purpose of this paper was to evaluate microtainer tubes for clinical chemistry and hematological analysis of serum.

Conclusion

Serum collected in Microtainer tubes was free of leukocyte and platelet contamination and had no visible hemolysis. Sodium levels were lower in specimens collected in Microtainer tubes, but the difference was not clinically significant. There was no effect of storage in the Microtainer tube for up to 24 h on potassium and lactate dehydrogenase levels

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube type, and storage on clinical chemistry and hematological parameters in serum from pediatric patients.

Summary of Findings

Serum collected in Microtainer tubes was free of leukocyte and platelet contamination and had no visible hemolysis. No effect of collection tube type was found on cholesterol, triglycerides, alkaline phosphatase, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, total protein, creatinine, sodium, potassium, chloride, calcium, thyroxine, billirubin, blood urea nitrogen or phosphorus levels. Sodium levels were lower in specimens collected in Microtainer tubes, but the difference was not clinically significant. There was no effect of storage in the Microtainer tube for up to 24 h on potassium and lactate dehydrogenase levels.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Cell count/volumeLight microscopy
    ProteinClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    Electrolyte/MetalFlame emission photometry
    PeptideRadioimmunoassay
    ProteinHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    3 mL BD tube
    Microtainer tube
    StorageStorage duration
    0 h
    12 h
    5 h