Evaluation of a new blood-collecting device ("microtainer") that is suited for pediatric use.
Hicks JM, Rowland GL, Buffone GJ
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type, and storage on clinical chemistry and hematological parameters in serum from pediatric patients.
Summary of Findings
Serum collected in Microtainer tubes was free of leukocyte and platelet contamination and had no visible hemolysis. No effect of collection tube type was found on cholesterol, triglycerides, alkaline phosphatase, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, creatine kinase, total protein, creatinine, sodium, potassium, chloride, calcium, thyroxine, billirubin, blood urea nitrogen or phosphorus levels. Sodium levels were lower in specimens collected in Microtainer tubes, but the difference was not clinically significant. There was no effect of storage in the Microtainer tube for up to 24 h on potassium and lactate dehydrogenase levels.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Cell count/volumeLight microscopyProteinClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerGasClinical chemistry/auto analyzerElectrolyte/MetalFlame emission photometryPeptideRadioimmunoassayProteinHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution 3 mL BD tubeMicrotainer tubeStorage Storage duration 0 h12 h5 h