The effects of heparin anticoagulants and fill volume in blood gas syringes on ionized calcium and magnesium measurements.

Toffaletti JG, Wildermann RF

Publication: Clin Chim Acta (2001) • Vol 304 • Page 147-51PMID: 11165211View on PubMed

Purpose

The purpose of this paper was to determine the effects of fill volume and heparin formulation on electrolyte levels in blood.

Conclusion

Compared to tubes without anticoagulant, serum and plasma specimens in syringes containing heparin in an inert web or as a blend of lithium and zinc heparin had lower calcium levels, particularly when the syringe was filled only halfway. No difference in calcium levels was found in blood from syringes containing electrolyte balanced heparin at 27 IU/mL or 40 IU/mL. When fill volume was halved, magnesium levels were lower in syringes containing electrolyte balanced heparin and higher in syringes with heparin in an inert web or in serum separator tubes (SST) compared to blood without anticoagulant. Specimens in completely filled syringes and those containing lithium and zinc heparin had magnesium levels similar to specimens collected without anticoagulant.

Studies (1)

Purpose

The purpose of this study was to determine the effects of fill volume and heparin formulation on calcium and magnesium levels in serum and plasma from healthy donors. Blood was collected and immediately transferred to evacuated blood collection tubes or syringes containing anticoagulant. Initial calcium and magnesium levels were lowered or raised in select specimens to broaden the concentration range.

Summary of Findings

Compared to tubes without anticoagulant, serum and plasma specimens in syringes containing heparin in an inert web or as a blend of lithium and zinc heparin had lower calcium levels, particularly when the syringe was filled only halfway. No difference in calcium levels was found in blood from syringes containing electrolyte balanced heparin at 27 IU/mL or 40 IU/mL. When fill volume was halved, magnesium levels were lower in syringes containing electrolyte balanced heparin and higher in syringes with heparin in an inert web or in SST compared to blood without anticoagulant. Specimens in completely filled syringes and those containing lithium and zinc heparin had magnesium levels similar to specimens collected without anticoagulant.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    0.41-1.06 mmol/L magnesium
    0.94-2.09 mmol/L calcium
    Biospecimen AcquisitionAnticoagulant
    Electrolyte balanced heparin
    Heparin
    Lithium heparin
    None
    Zinc heparin
    Biospecimen AcquisitionType of collection container/solution
    AVL COMBI syringe
    Marquest GasLyte syringe
    Martell Pulset syringe
    Plain tube
    Radiometer PICO syringe
    Serum separator tube(SST)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    Full
    Half-full
    Biospecimen Aliquots and ComponentsBiospecimen components
    27 IU/mL electrolyte balanced heparin
    40 IU/mL electrolyte balanced heparin
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Whole blood