Storage of specimens in "Microtainer Amber Tubes" protects bilirubin from degradation by light.
Desjardins PR, Mactavish G, Schoemperlen J
Purpose
The purpose of this paper was to determine if Microtainer amber tubes prevent degradation of bilirubin in serum.
Conclusion
Bilirubin concentrations remained constant in serum stored in Microtainer amber tubes, regardless of assay. Concentration and assay-specific effects were noted in specimens stored in translucent Microtainer tubes.
Studies (1)
Purpose
The purpose of this study was to determine the effects of initial bilirubin concentration, assay type, and storage container and duration on the degradation of bilirubin in serum. One serum pool was prepared for each bilirubin concentration.
Summary of Findings
Bilirubin concentrations remained constant in serum stored in Microtainer amber tubes, regardless of assay. In contrast, in serum with a low initial concentration, bilirubin levels declined with storage in translucent Microtainer tubes when assayed by the Diazo method, but remained unaffected by storage when assayed by the spectrophotometric method. However, when initial bilirubin levels were high, a decline with storage in translucent Microtainer tubes was observed with the spectrophotometric method and not the Diazo method.
Biospecimens
FluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Small moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 20-30 umol/L bilirubin230-260 umol/L bilirubinStorage Storage conditions Well lit roomStorage Time at room temperature 0 h1 h2 h27 h3 h4 h5 h6 hStorage Type of storage container BD Microtainer amber tubesBD translucent Microtainer tubeClinical chemistry/auto analyzer Specific Detection method Diazo methodSpectrophotometric method