Skip to main content

Absorption of therapeutic drugs by barrier gels in serum separator blood collection tubes. Volume- and time-dependent reduction in total and free drug concentrations.

Dasgupta A, Dean R, Saldana S, Kinnaman G, McLawhon RW

Publication: Am J Clin Pathol (1994) • Vol 101 • Page 456-61PMID: 8160636View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage in a serum separator tube (SST) on concentrations of therapeutic drugs in serum.

Conclusion

Significant declines in the concentrations of phenytoin, phenobarbital, carbamazepine, quinidine, and lidocaine, but not salicylate or theophylline occurred within 20 min when serum was stored in Vacutainer SST. In contrast, when serum was stored in Corvac SST, only lidocaine concentrations decreased, and when serum was stored in red-top Vacutainer tubes, no significant changes in any therapeutic drug concentrations were found. The effects of storage in Vacutainer SST and Corvac SST were most dramatic when the specimen volume was small and when the specimens were stored for 48 h. The presence of protein carriers in normal serum was found to partially attenuate the effects of storage. The authors concluded that the barrier gel present in Vacutainer SST rapidly adsorbs therapeutic drugs during storage and that the effect was greatest when the volume was small or when a serum ultrafiltrate was stored.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage in SST and specimen volume on concentrations of therapeutic drugs in serum. Patient serum and therapeutic drug-spiked serum were used.

Summary of Findings

Significant declines in the concentrations of phenytoin, phenobarbital, carbamazepine, quinidine, and lidocaine, but not salicylate or theophylline occurred within 20 min when serum was stored in Vacutainer SST. In contrast, when serum was stored in Corvac SST, only lidocaine concentrations decreased, and when serum was stored in red-top Vacutainer tubes, no significant changes in any therapeutic drug concentrations were found. The effects of storage in Vacutainer SST and Corvac SST were most dramatic when the specimen volume was small and when the specimens were stored for 48 h. The presence of protein carriers in normal serum was found to partially attenuate the effects of storage. The authors concluded that the barrier gel present in Vacutainer SST rapidly adsorbs therapeutic drugs during storage and that the effect was greatest when the volume was small or when a serum ultrafiltrate was stored.
  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Normal
  • Platforms

    Small moleculeImmunoassay
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    1000 ul
    1500 ul
    200 ul
    300 ul
    500 ul
    750 ul
    Biospecimen Aliquots and ComponentsFiltration
    Normal serum
    Ultrafiltrated serum
    StorageStorage duration
    0 min
    1 h
    2 h
    20 min
    24 h
    40 min
    48 h
    6 h
    StorageStorage temperature
    4 degrees C
    StorageType of storage container
    Corvac SST
    Red-top Vacutainer
    Vacutainer SST