Skip to main content

Recommended "order of draw" for collecting blood specimens into additive-containing tubes.

Calam RR, Cooper MH

Publication: Clin Chem (1982) • Vol 28 • Page 1399PMID: 7074955View on PubMed

Purpose

The purpose of this paper was to assess the effects of anticoagulant and draw order on calcium and potassium levels in blood.

Conclusion

When blood was collected first into the K3EDTA tube, levels of potassium were increased and calcium levels were decreased in the second collection which was drawn into a serum separator tube (SST), but when the SST was used for the first collection and the K3EDTA tube was used for the second, no interference was observed.

Studies (1)

Purpose

The purpose of this study was to determine effects of anticoagulant and draw order on calcium and potassium levels in blood. Blood was collected from a single venipuncture, first into an SST, then a K3EDTA tube. After a second venipuncture in the same individual, the tube order used for collection of two specimens was reversed.

Summary of Findings

When blood was collected first into the K3EDTA tube, levels of potassium were increased and calcium levels were decreased in the second collection which was drawn into an SST, but when the SST was used for the first collection and the K3EDTA tube was used for the second, no interference was observed. The authors note that similar interference can occur with other anticoagulants so a strict order of draw should be followed.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot sequential collection
    1st collection
    2nd collection
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum