Interchangeability of B type natriuretic peptide measurements in lithium heparin and K2EDTA: agreement between matrices and differences in diagnostic accuracy.
Stenner E, Buiatti A, Bussani A, Merlo M, d'Amario C, Barbati G, Sinagra G, Biasioli B
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of anticoagulant on BNP measurements in plasma.
Summary of Findings
A linear relationship was noted between BNP tests using EDTA and lithium heparin. At a threshold of 100 ng/L, the sensitivity and specificity were 1 and 0.558 for EDTA-plasma and 0.92 and 0.63 for heparin-plasma, respectively. Results suggest that lithiuim heparin can be used as an alternative to the recommended EDTA for BNP assays. The authors note the small sample size of specimens collected with both anticoagulants (52) as opposed to those collected with EDTA and compared to computer simulated heparin (239) is a weakness of this study.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Cardiovascular DiseaseOther diagnosesPlatforms
PeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Symptoms of heart failureBiospecimen Acquisition Anticoagulant Lithium heparinPotassium EDTA