Therapeutic monitoring of cyclosporine following pediatric bone marrow transplantation: problems with sampling from silicone central venous lines.
Leson CL, Bryson SM, Giesbrecht EE, Saunders EF
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection method on cyclosporine concentrations in pediatric leukemia and blackfan diamond anemia patients.
Summary of Findings
When cyclosporine was administered through iv line, blood obtained through CVL had 1.3-2.9 fold higher cyclosporine levels than peripheral capillary specimens, however, specimen collection method had no effect on the measurement of cyclosporine when the antibiotic was administered orally. The authors conclude that there is a problem with adsorption of cyclosporine by the silicone in the CVL during drug administration.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NeoplasticLeukemiaNeoplasticPediatricOther diagnosesPlatforms
Small moleculeHPLCPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Blackfan diamond anemiaPreaquisition Other drugs IV cyclosporineOral cyclosporineBiospecimen Acquisition Anatomical location of blood draw Central veinPeripheral capillary