Modification of the EMIT immunoassay for the measurement of unbound mycophenolic acid in plasma.

Rebollo N, Calvo MV, Martín-Suárez A, Domínguez-Gil A

Publication: Clin Biochem (2011) • Vol 44 • Page 260-3PMID: 20923675View on PubMed

Purpose

The purpose of this paper was to validate a modified immunoassay for the analysis of free mycophenolic acid (fMPA) in ultrafiltrated plasma and determine the effects of ultrafiltration conditions.

Conclusion

The fMPA concentrations corresponded to the expected values when specimens were diluted with MPA-free calibrator 1, but when they were diluted with MPA-free ultrafiltered plasma, fMPA concentrations were higher than expected. Concentrations of fMPA in ultrafilered plasma were not affected by specimen volume, or centrifugation speed or duration, but significantly lower concentrations of fMPA were found when plasma ultrafiltration occurred at 23-26 degrees C instead of 37 degrees C.

Studies (1)

Purpose

The purpose of this study was to determine the effects of specimen dilution, ultrafiltration temperature, volume, and centrifugation speed (1000 x g versus 2000 x g) and duration (10, 20 30, 40, or 60 min) on levels of fMPA in ultrafiltered plasma.

Summary of Findings

The fMPA concentrations corresponded to the expected values when specimens were diluted with MPA-free calibrator 1, but when they were diluted with MPA-free ultrafiltered plasma, fMPA concentrations were higher than expected. The authors report that the effect of dilution can be easily corrected mathematically. Concentrations of fMPA in ultrafilered plasma were not affected by specimen volume, or centrifugation speed or duration, but significantly lower concentrations of fMPA were found when plasma ultrafiltration occurred at 23-26 degrees C instead of 37 degrees C (p<0.05).

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Other diagnoses
  • Platforms

    Small moleculeImmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionDiagnosis/ patient condition
    Post-transplant renal allograft
    Biospecimen Aliquots and ComponentsAliquot size/volume
    0.5 mL
    1 mL
    Biospecimen Aliquots and ComponentsBiospecimen components
    Diluted with MPA-free calibrator 1
    Diluted with MPA-free plasma
    Biospecimen Aliquots and ComponentsCentrifugation
    Multiple durations compared
    Multiple speeds compared
    Biospecimen Aliquots and ComponentsFiltration
    Ultrafiltration at 23-26 degrees C