Modification of the EMIT immunoassay for the measurement of unbound mycophenolic acid in plasma.
Rebollo N, Calvo MV, Martín-Suárez A, Domínguez-Gil A
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of specimen dilution, ultrafiltration temperature, volume, and centrifugation speed (1000 x g versus 2000 x g) and duration (10, 20 30, 40, or 60 min) on levels of fMPA in ultrafiltered plasma.
Summary of Findings
The fMPA concentrations corresponded to the expected values when specimens were diluted with MPA-free calibrator 1, but when they were diluted with MPA-free ultrafiltered plasma, fMPA concentrations were higher than expected. The authors report that the effect of dilution can be easily corrected mathematically. Concentrations of fMPA in ultrafilered plasma were not affected by specimen volume, or centrifugation speed or duration, but significantly lower concentrations of fMPA were found when plasma ultrafiltration occurred at 23-26 degrees C instead of 37 degrees C (p<0.05).
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedOther diagnosesPlatforms
Small moleculeImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition Post-transplant renal allograftBiospecimen Aliquots and Components Aliquot size/volume 0.5 mL1 mLBiospecimen Aliquots and Components Biospecimen components Diluted with MPA-free calibrator 1Diluted with MPA-free plasmaBiospecimen Aliquots and Components Centrifugation Multiple durations comparedMultiple speeds comparedBiospecimen Aliquots and Components Filtration Ultrafiltration at 23-26 degrees CUltrafiltration at 37 degrees C