Prevention of preanalytical false-positive increases of cardiac troponin I on the Unicel DxI 800 analyzer.

Hejl CG, Astier HT, Ramirez JM

Publication: Clin Chem Lab Med (2008) • Vol 46 • Page 1789-90PMID: 19055457View on PubMed

Purpose

The purpose of this paper was to determine if faster centrifugation could prevent false positives in the cardiac troponin I assay.

Conclusion

When blood was centrifuged at 2300 x g for 15 minutes, 4% of specimens collected in glass tubes and 3% of specimens collected in plastic tubes gave false positives in the cardiac troponin I assay. Centrifugation of specimens for 15 min at 3000 x g did not decrease false positives, but ultracentrifugation at 6700 x g for 5 minutes reduced the false positive rate to 1.1%.

Studies (1)

Purpose

The purpose of this study was to determine the effects of blood centrifugation speed on cardiac troponin I quantification in plasma.

Summary of Findings

When blood was centrifuged at 2300 x g for 15 minutes, 4% of specimens collected in glass tubes and 3% of specimens collected in plastic tubes gave physician identified false positives in the cardiac troponin I assay. Centrifugation of specimens for 15 min at 3000 x g did not decrease false positives, but ultracentrifugation at 6700 x g for 5 minutes reduced the false positive rate to 1.1%.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Cardiovascular Disease
    Coronary Artery Disease
    Not specified
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Glass tube
    Plastic tube
    Biospecimen Aliquots and ComponentsCentrifugation
    Multiple durations compared
    Multiple speeds compared