Prevention of preanalytical false-positive increases of cardiac troponin I on the Unicel DxI 800 analyzer.
Hejl CG, Astier HT, Ramirez JM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood centrifugation speed on cardiac troponin I quantification in plasma.
Summary of Findings
When blood was centrifuged at 2300 x g for 15 minutes, 4% of specimens collected in glass tubes and 3% of specimens collected in plastic tubes gave physician identified false positives in the cardiac troponin I assay. Centrifugation of specimens for 15 min at 3000 x g did not decrease false positives, but ultracentrifugation at 6700 x g for 5 minutes reduced the false positive rate to 1.1%.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
Cardiovascular DiseaseCoronary Artery DiseaseNot specifiedPlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Glass tubePlastic tubeBiospecimen Aliquots and Components Centrifugation Multiple durations comparedMultiple speeds compared