Heparin interference in whole blood sodium measurements in a pediatric setting.
Yip PM, Chan MK, Zielinski N, Adeli K
Purpose
Conclusion
Studies (2)
Purpose
The purpose of this study was to determine the effects of excess heparin in partially filled syringes on sodium levels in whole blood and plasma. Blood was collected in plain glass tubes without anticoagulant, aliquoted into syringes containing anticoagulant and then analyzed within 30 minutes.
Summary of Findings
When heparin containing syringes were filled completely, there were nonsignificant reductions in sodium concentration in whole blood compared to levels without anticoagulant. Partial filling of syringes had no effect on sodium concentrations when Gaslyte syringes containing reduced heparin were used. However, filling PICO50 syringes containing electrolyte balanced heparin to 0.5 mL or Aspirator syringes containing standard heparin to 1 mL or 0.5 mL resulted in decreased measured sodium levels. When heparin was added to leftover patient blood, there were significant decreases in sodium concentration in whole blood, but not in plasma or serum.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant Calcium-titrated sodium heparinHeparinLithium heparinNoneBiospecimen Acquisition Type of collection container/solution Aspirator syringeGaslyte syringePICO50 syringeBiospecimen Aliquots and Components Aliquot size/volume 0.5 mL1.0 mL2.0 mL3.0 mLBiospecimen Aliquots and Components Blood and blood products PlasmaSerumWhole blood
Purpose
The purpose of this study was to determine the effects of collection container and sodium concentration on sodium quantification in pediatric patients.
Summary of Findings
There was a strong correlation between sodium levels in blood collected in capillary tubes containing heparin and serum collected in microtainer tubes (r=0.68), but there was a negative bias for specimens collected in capillary tubes. The bias in sodium levels was present for the entire range of values, but was strongest when sodium concentrations were above 140 mmol/L.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 125-150 mmol/L sodiumPreaquisition Patient age NeonatesBiospecimen Acquisition Anticoagulant HeparinNoneBiospecimen Acquisition Type of collection container/solution Capillary tubeMicrotainer clotBiospecimen Aliquots and Components Blood and blood products SerumWhole blood