A comparison of hard and soft glass blood-drawing tubes.
Laessig RH, Hassemer DJ, Hoffman GL, Westgard JO, Carey RN, Feldbruegge DH
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube type and storage on serum clinical chemistry analytes.
Summary of Findings
The use of hard or soft glass disposable evacuated blood-drawing tubes had no clinically significant effects on 34 clinical chemistry analytes using 66 different methods of analysis. However, in four cases, small but statistically significant differences between the tube types were observed including differences in triglycerides, unsaturated iron-binding capacity, uric acid, and inorganic phosphorus. Repeat analysis of uric acid and inorganic phosphorus by other methods did not detect differences. In addition, there were no clinically or statistically significant differences in analyte levels between specimens refrigerated for 72 h in the two types of glass tubes.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerCarbohydrateClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerElectrolyte/MetalFlame emission photometryElectrolyte/MetalColorimetric assayPeptideRadioimmunoassaySmall moleculeRadioimmunoassaySteroidRadioimmunoassayGlycoproteinRadial immunodiffusion assayProteinRadial immunodiffusion assayElectrolyte/MetalAtomic absorption spectroscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Hard glass evacuated blood-drawing tubesSoft glass evacuated blood-drawing tubesStorage Storage duration 72 hLess than 60 minStorage Storage temperature RefrigeratedClinical chemistry/auto analyzer Specific Technology platform AutoAnalyzer IBASICDuPont ACASMAC