False increases of troponin I attributable to incomplete separation of serum.
Nosanchuk JS, Combs B, Abbott G
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of recentrifugation and collection of plasma instead of serum on troponin I levels using the AxSYM assay.
Summary of Findings
Initial levels of Troponin I were higher in serum than in plasma, however, after serum was centrifuged a second time, the troponin I levels were the same as in plasma. The authors believe the antibody used in the AxSYM assay may interact with fibrin resulting in falsely high Troponin I levels when serum separation is incomplete.
Biospecimens
FluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Blood and blood products PlasmaSerumBiospecimen Aliquots and Components Centrifugation Different number of centrifugation steps compared