False increases of troponin I attributable to incomplete separation of serum.

Nosanchuk JS, Combs B, Abbott G

Publication: Clin Chem (1999) • Vol 45 • Page 714PMID: 10222366View on PubMed

Purpose

The purpose of this paper was to determine the effects of recentrifugation and collection of plasma instead of serum on troponin I levels using the AxSYM assay.

Conclusion

Initial levels of Troponin I were higher in serum than in plasma, however, after serum was centrifuged a second time, the troponin I levels were the same as in plasma.

Studies (1)

Purpose

The purpose of this study was to determine the effects of recentrifugation and collection of plasma instead of serum on troponin I levels using the AxSYM assay.

Summary of Findings

Initial levels of Troponin I were higher in serum than in plasma, however, after serum was centrifuged a second time, the troponin I levels were the same as in plasma. The authors believe the antibody used in the AxSYM assay may interact with fibrin resulting in falsely high Troponin I levels when serum separation is incomplete.

  • Biospecimens

    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Different number of centrifugation steps compared