Assessment of the serum separator tube as an intermediate storage device within the laboratory.

Laessig RH, Hassemer DJ, Westgard JO, Carey RN, Feldbruegge DH, Schwartz TH

Publication: Am J Clin Pathol (1976) • Vol 66 • Page 653-7PMID: 970366View on PubMed

Purpose

The purpose of this paper was to determine the effects of storage of serum in serum separator tubes (SST) on clinical chemistry analyses.

Conclusion

Storage of serum for 48 hours in the SST in which specimens were collected and centrifuged had no effects on 86 clinical chemistry determinations when compared with serum that was transferred to a disposable closed culture tube for storage.

Studies (1)

Purpose

The purpose of this study was to determine the effects of refrigerated storage of serum in SST on clinical chemistry analyses.

Summary of Findings

Storage of serum for 48 hours in the SST in which specimens were collected and centrifuged had no clinically significant effects on 86 clinical chemistry determinations when compared with serum that was transferred to a disposable closed culture tube for storage. However, six small but statistically significant (p=0.05) changes were noted including changes in lactate dehydrogenase by SMAC, glucose by SMA 6/60 and AutoAnalyzer I, chloride by colorimetric determination, and lactic acid and lactate dehydrogenase by Dupont ACA. Notably, other methods of detecting these same analytes did not find any statistically significant changes.

  • Biospecimens

    FluidSerum
  • Preservation Types

    Other Preservative
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    CarbohydrateClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    ProteinElectrophoresis
    LipoproteinElectrophoresis
    Electrolyte/MetalFlame emission photometry
    Electrolyte/MetalColorimetric assay
    Small moleculeRadioimmunoassay
    SteroidRadioimmunoassay
    ProteinRadial immunodiffusion assay
    Electrolyte/MetalAtomic absorption spectroscopy
    Small moleculeGC
    PeptideRadioimmunoassay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageType of storage container
    Closed culture tube
    Serum separator tube
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Auto-Analyzer I
    Beckman Microzone System
    Dupont ACA
    SMA 12/60
    SMA 6/60
    SMAC