Comparison of single and repeat centrifugation of blood specimens collected in BD evacuated blood collection tubes containing a clot activator for cardiac troponin I assay on the ACCESS analyzer.
Beyne P, Vigier JP, Bourgoin P, Vidaud M
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection container and repeat centrifugation on analysis of cTnI in patients with chest pain or acute coronary syndrome.
Summary of Findings
When blood was collected in BD Thrombin tubes and centrifuged once, there was a positive bias in cTnI levels compared to blood collected in BD SST and centrifuged twice (p<0.0001). Importantly, the positive bias resulted in 6 false positives out of 64 specimens. When specimens collected in BD Thrombin tubes were centrifuged twice, the effect of collection container disappeared.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedCoronary Artery DiseaseCardiovascular DiseasePlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution BD Thrombin tubeBD Vacutainer Hemogard SSTBiospecimen Aliquots and Components Centrifugation Different number of centrifugation steps compared