Ethylbenzene and xylene from Sarstedt Monovette serum gel blood-collection tubes.
Streete PJ, Flanagan RJ
Purpose
The purpose of this paper was to determine the source of ethylbenzene and xylene contamination of whole blood.
Conclusion
Storage of blood at room temperature for 96 h in Monovette tubes resulted in measureable levels of o-xylene, ethylbenzene, and m-xylene. The magnitude of the increase was dependent on the age of the tubes with tubes furthest from expiration having the largest effect.
Studies (1)
Purpose
The purpose of this study was to determine the effects of storing blood in Starstedt Monovette tubes on the quantification of ethylbenzene and xylene. Expired volatile-free hemolyzed blood-bank whole blood was used for all assays.
Summary of Findings
Storage of blood at room temperature for 96 h in Monovette tubes resulted in measureable levels of o-xylene, ethylbenzene, and m-xylene. The magnitude of the increase was dependent on the age of the tubes with tubes furthest from expiration having the largest effect. The authors report that Starstedt Monovette serum gel tubes, and Becton Dickinson Vacutainer STT gel and clot activator tubes do not cause an increase in the concentration of volatiles.
Biospecimens
FluidBloodPreservation Types
FrozenDiagnoses
NormalPlatforms
Small moleculeHeadspace gas chromatographyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Time at room temperature 120 h24 h96 hStorage Type of storage container Starstedt Monovette serum gel tube