Unusual interference from primary collection tube in a high-performance liquid chromatography assay of amiodarone.
Murthy VV
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection tube on quantification of AMIO and DESAMIO in protein depleted serum from cardiovascular disease patients.
Summary of Findings
When serum was collected in Corvac Monoject, but not Becton Dickinson Vacutainer tubes, an extraneous peak near the expected peak for DESAMIO was observed. The peak was present in serum regardless of whether the serum was stored in the Monoject tube or immediately transferred to a glass tube.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Cardiovascular DiseasePlatforms
Small moleculeHPLCPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Becton Dickinson Vacutainer tubesCorvac Monoject tubesStorage Type of storage container Corvac Monoject tubesGlass tubes