Skip to main content

Unusual interference from primary collection tube in a high-performance liquid chromatography assay of amiodarone.

Murthy VV

Publication: J Clin Lab Anal (1997) • Vol 11 • Page 232-4PMID: 9219066View on PubMed

Purpose

The purpose of this paper was to determine the source of interference with the HPLC assay of the drug amiodarone (AMIO) and its therapeutic metabolite desethylamiodarone (DESAMIO) in serum.

Conclusion

When serum was collected in Corvac Monoject, but not Becton Dickinson Vacutainer tubes, an extraneous peak near the expected peak for DESAMIO was observed. The peak was present in serum regardless of whether the serum was stored in the Monoject tube or immediately transferred to a glass tube.

Studies (1)

Purpose

The purpose of this study was to determine the effects of collection tube on quantification of AMIO and DESAMIO in protein depleted serum from cardiovascular disease patients.

Summary of Findings

When serum was collected in Corvac Monoject, but not Becton Dickinson Vacutainer tubes, an extraneous peak near the expected peak for DESAMIO was observed. The peak was present in serum regardless of whether the serum was stored in the Monoject tube or immediately transferred to a glass tube.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Cardiovascular Disease
  • Platforms

    Small moleculeHPLC
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Becton Dickinson Vacutainer tubes
    Corvac Monoject tubes
    StorageType of storage container
    Corvac Monoject tubes
    Glass tubes