Spuriously low plasma propranolol concentrations resulting from blood collection methods.
Cotham RH, Shand D
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood collection container and method on plasma and red blood cell propranolol concentration. Spiked normal blood and patient blood were included in this study.
Summary of Findings
Compared to specimens in glass tubes, spiked blood specimens collected in purple, red or green Vacutainer tubes, but not blue Vacutainer tubes, plastic syringes, or any tested Venoject tubes, had lower plasma propranolol levels. Additionally, purple, red, and green Vacutainer tubes had higher propranolol levels in the red blood cell fraction. When the green stopper used on Vacutainer tubes was replaced with glass or parafilm, the effects of collection container on propranolol levels vanished. A reduction in plasma propranolol levels was observed when specimens were collected by indwelling catheter through polyvinyl tubing rather than by venipuncture.
Biospecimens
FluidBloodFluidPlasmaCellRed Blood CellsPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
Small moleculeFluorometrySmall moleculeEquilibrium dialysisPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Indwelling catheterVenipunctureBiospecimen Acquisition Type of collection container/solution Blue-stoppered Venoject tubeBue-stoppered Vacutainer tubeGlass-stoppered tubeGreen-stoppered Vacutainer tubeGreen-stoppered Venoject tubePlastic syringePurple-stoppered Vacutainer tubeRed-stoppered Vacutainer tubeRed-stoppered Venoject tubeBiospecimen Aliquots and Components Blood and blood products PlasmaRed blood cellsWhole blood