Differences between values for plasma and serum in tests performed in the Ektachem 700 XR Analyzer, and evaluation of "plasma separator tubes (PST)".
Doumas BT, Hause LL, Simuncak DM, Breitenfeld D
Purpose
Conclusion
Studies (3)
Purpose
The purpose of this study was to compare levels of 25 clinical chemistry analytes in plasma and serum specimens.
Summary of Findings
No differences were found between plasma and serum levels of creatinine, glucose, alkaline phosphatase (AP), HCO3, calcium, chloride, creatinine kinase (CK), glucose, gamma-glutamyltransferase (gamma-GT), LD, magnesium (Mg), triglycerides, urea, nitrogen and uric acid. Higher levels of total cholesterol, high density lipoprotein (HDL) cholesterol, total bilirubin, and sodium (Na) were found in plasma than in serum, and higher levels of lipase and alanine aminotransferase (ALT) were found in serum than plasma, but these differences were not enough to be clinically relevant. Plasma levels of phosphorous and K were lower than in serum, and plasma levels of total protein, amylase and AST were higher than in serum.
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
Electrolyte/MetalClinical chemistry/auto analyzerCarbohydrateClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant HeparinNoneBiospecimen Aliquots and Components Blood and blood products PlasmaSerum
Purpose
The purpose of this study was to determine the effect of collection of specimens in PST rather than heparinized tubes on 25 clinical chemistry analytes in plasma.
Summary of Findings
Collection of plasma in PST rather than heparinized tubes resulted in clinically insignificant changes of less than 4% in the levels of all analytes.
Biospecimens
FluidPlasmaFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
CarbohydrateClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Heparinized tubePlasma separation tube
Purpose
Summary of Findings
Biospecimens
FluidPlasmaPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
ProteinClinical chemistry/auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerCarbohydrateClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 2 days2 h24 h3 days4 days4 h5 days6 h8 h< 30 minStorage Storage temperature 4 degrees CRoom temperatureStorage Type of storage container Plasma separation tube