Glycerol lubricant: potential source of error in commercially prepared blood specimen tubes.

Baum D

Publication: Clin Chem (1968) • Vol 14 • Page 73-4PMID: 5634642View on PubMed

Purpose

The purpose of this paper was to determine the source of elevated glycerol levels in blood specimens.

Conclusion

Blood specimens stored in Vacutainer tubes showed elevated levels of glycerol compared to specimens stored in washed tubes.

Studies (1)

Purpose

The purpose of this study was to determine the effects of storage container on blood glycerol levels.

Summary of Findings

Blood specimens stored in Vacutainer tubes showed elevated levels of glycerol compared to specimen stored in washed tubes. The authors report that the manufacturer identified the source of glycerol as stopper lubricant.
  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Small moleculeEnzyme assay
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageType of storage container
    Laboratory washed tube
    Vacutainer tube