Trace metal determinations in blood obtained in evacuated collection tubes.
Williams DM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of collection container tube on copper, lead and zinc in serum and plasma from normal volunteers and cancer patients.
Summary of Findings
The type of collection container did not affect copper or iron levels in serum or plasma. In contrast, zinc levels were as much as 2.5 fold higher when specimens were collected in Vacutainer tubes rather than glass. However, zinc levels in serum collected in the "metal free" Vacutainers were only 10% higher (p<0.01) than specimens collected in glass tubes, and there was a strong correlation between the values (r=0.973).
Biospecimens
FluidPlasmaFluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
NormalNeoplasticPlatforms
Electrolyte/MetalAtomic absorption spectroscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Acid-washed glassHeparinized (green) VacutainerLead free (amber) VacutainerMetal-free (blue) VacutainerPlain (red) VacutainerBiospecimen Aliquots and Components Blood and blood products PlasmaSerum