Effect of magnesium contamination in evacuated blood collection tubes on the prothrombin time test and ISI calibration using recombinant human thromboplastin and different types of coagulometer.

van den Besselaar AM, Rutten WP, Witteveen E

Publication: Thromb Res (2005) • Vol 115 • Page 239-44PMID: 15617747View on PubMed

Purpose

The purpose of this paper was to determine the effects of blood collection tube and coagulometer on prothrombin time (PT) and international sensitivity index (ISI).

Conclusion

Regardless of coagulometer used, specimens collected in Vacutainer tubes or in the presence of added Mg had a significantly shorter mean PT compared to specimens collected in Venoject II tubes (p<0.01). Further, the PT ratio was lower and the ISI was higher in specimens collected in Vacutainer tubes rather than Venoject II tubes, but the magnitude of the change depended on the coagulometer.

Studies (2)

Purpose

The purpose of this study was to determine the effects of blood collection tube and coagulometer on PT and ISI from normal volunteers and patients on oral anticoagulants. Plasma was kept at room or ambient temperature during transport prior to analysis.

Summary of Findings

Regardless of coagulometer used, specimens collected in Vacutainer tubes had a significantly shorter mean PT compared to specimens collected in Venoject II tubes (p<0.01) which the authors attribute to the presence of Mg in Vacutainer tubes but not Venoject II tubes. The PT ratio was lower and the ISI was higher in specimens collected in Vacutainer tubes rather than Venoject II tubes. The magnitude of the change in ISI was greatest with the Electra 1000 (MLA) coagulometer and least with the ACL Advance coagulometer.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Normal
  • Platforms

    MorphologyHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionOther drugs
    None
    Oral anticoagulants
    Biospecimen AcquisitionType of collection container/solution
    Vacutainer
    Venoject II
    Hematology/ auto analyzer SpecificTechnology platform
    ACL 300 coagulometer
    ACL Advance coagulometer
    Electra 1000 (MLA) coagulometer