Prothrombin time ratio is reduced by magnesium contamination in evacuated blood collection tubes.
van den Besselaar AM, van Dam W, Sturk A, Bertina RM
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of Mg contamination on PT in blood and plasma specimens from patients on oral anticoagulants and normal controls. Plasma was dialyzed to remove citrate and endogenous ions prior to experimental addition of citrate and magnesium.
Summary of Findings
At concentrations found in citrate Vacutainer tubes and stoppers, Mg significantly decreased PT in normal plasma and in plasma from patients on oral anticoagulants. However, the effect of Mg was greater in plasma from patients on oral anticoagulants and when plasma contained citrate. Because the effect of Mg was greater in patients on oral anticoagulants, increasing Mg concentration resulted in decreased PT-ratios and INR. When blood specimens were collected in freshly prepared citrate without magnesium, or in citrate with magnesium added to mimic levels in Vacutainer tubes, a 0.9-6.6% decrease in INR was found.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
FrozenDiagnoses
Not specifiedNormalPlatforms
MorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 0.2 mmol/L Mg0.5 mmol/L Mg1.0 mmol/L Mg2.0 mmol/L MgPreaquisition Other drugs NoneOral anticoagulantsBiospecimen Acquisition Anticoagulant NoneSodium citrateBiospecimen Aliquots and Components Blood and blood products PlasmaWhole blood