Importance of blood-collection tubes in plasma lidocaine determinations.
Stargel WW, Roe CR, Routledge PA, Shand DG
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of blood or plasma contact with collection tube stoppers on lidocaine concentration. Clinical specimens and lidocaine spiked specimens from a normal volunteer were collected in plastic syringes and whole blood or plasma was then distributed to the various tubes.
Summary of Findings
In 25 clinical specimens, contact of whole blood with the red stoppers on Vacutainer tubes resulted in a decrease in mean lidocaine concentration in serum from 6.5 ug/mL to 4.9 ug/mL (p<0.001). Similarly, there was a significant decrease in measured lidocaine when whole blood, but not plasma, was added to red or green Vacutainer or red Monoject tubes containing exogenous lidocaine. The effect of using plastic Vacutainer tubes versus an all glass system decreased as lidocaine concentrations increased. However, in plastic Vacutainer tubes or an all glass system, increasing lidocaine concentration decreased the ratio of lidocaine in blood to lidocaine in plasma. Finally, when blood was collected in Vacutainer tubes, lidocaine binding of plasma was 28%, versus 56% when blood was collected in glass.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedNormalPlatforms
Small moleculeImmunoassayPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 10 ug/mL lidocaine12 ug/mL lidocaine2 ug/mL lidocaine4 ug/mL lidocaine6 ug/mL lidocaine8 ug/mL lidocaineBiospecimen Aliquots and Components Blood and blood products PlasmaSerumWhole bloodStorage Storage conditions In contact with stopper or with inversionWithout stopperStorage Type of storage container All glass systemGreen VacutainerRed MonojectRed Vacutainer