Influence of tube type on the antiglobulin test.
Black D, Kay J
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of tube type, reagent source, centrifugation speed (400 x g versus 1000 x g) and antibody on AHG in red blood cells.
Summary of Findings
Generally the strongest AHG reactions were observed with the IMMUNO reagents and the weakest with the Blood Group Reference Laboratory (BGRL) reagents. Regardless of reagent source and antibody, AHG reactions increased with faster centrifugation and when the reaction occurred in glass rather than in plastic tubes. The decreased reaction in plastic tubes was found to be due to diminished anti-immunoglobulin G (anti-IgG) activity and not a change in anti-C3 activity. The effect was reduced as the concentration of antibody increased, but only occurred when both washing and the reaction were done in the plastic tube.
Biospecimens
CellRed Blood CellsPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
ProteinAntiglobulin testProteinMacroscopic observationPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Multiple speeds comparedStorage Type of storage container GlassPlasticAntiglobulin test Specific Antibody dilution/concentration 1:100 anti-IgG1:50 anti-IgG1:75 anti-IgG1:80 anti-IgGAntiglobulin test Specific Targeted peptide/protein Anti-C3Anti-DAnti-FyAnti-IgGAnti-JkAnti-KAnti-LeAnti-SAnti-cAnti-sAntiglobulin test Specific Technology platform Blood Group Reference LaboratoryIMMUNOScottish Antibody Production Unit