Interference caused by the contents of serum separator tubes in the Vitros CRP assay.
Chang CY, Lu JY, Chien TI, Kao JT, Lin MC, Shih PC, Yan SN
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of specimen volume, and collection or storage in SST tubes on measurement of CRP in serum from hospitalized patients with inflammatory disease.
Summary of Findings
When blood was collected in glass tubes rather than SST, lower levels of CRP were found in the resulting serum (p<0.0001). The difference between collection tubes increased as CRP concentration in the specimen increased, but specimen volume had no effect. When serum obtained in a glass tube was transferred to a SST and stored for 15 min, there was a similar increase in CRP values (p<0.0001) to that observed when specimens were collected in SST rather than glass tubes.
Biospecimens
FluidSerumFluidBloodPreservation Types
None (Fresh)Diagnoses
Other diagnosesPlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Biomarker level 14 mg/L CRP5 mg/L CRP53 mg/L CRPPreaquisition Diagnosis/ patient condition Hospitalized with inflammatory diseaseBiospecimen Acquisition Type of collection container/solution Glass tubeSerum separator tubeBiospecimen Aliquots and Components Aliquot size/volume 0.5 mL1 mL1.5 mL2 mLStorage Time at room temperature 0 min15 min