Skip to main content

Interference caused by the contents of serum separator tubes in the Vitros CRP assay.

Chang CY, Lu JY, Chien TI, Kao JT, Lin MC, Shih PC, Yan SN

Publication: Ann Clin Biochem (2003) • Vol 40 • Page 249-51PMID: 12803837View on PubMed

Purpose

The purpose of this paper was to determine the effects of collecting blood in serum separator tubes (SST) on the measurement of C-reactive protein (CRP) in serum.

Conclusion

Collection or storage of serum in SST instead of glass tubes led to an increase in measured CRP levels that was dependent on CRP concentration, but not specimen volume.

Studies (1)

Purpose

The purpose of this study was to determine the effects of specimen volume, and collection or storage in SST tubes on measurement of CRP in serum from hospitalized patients with inflammatory disease.

Summary of Findings

When blood was collected in glass tubes rather than SST, lower levels of CRP were found in the resulting serum (p<0.0001). The difference between collection tubes increased as CRP concentration in the specimen increased, but specimen volume had no effect. When serum obtained in a glass tube was transferred to a SST and stored for 15 min, there was a similar increase in CRP values (p<0.0001) to that observed when specimens were collected in SST rather than glass tubes.

  • Biospecimens

    FluidSerum
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Other diagnoses
  • Platforms

    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    PreaquisitionBiomarker level
    14 mg/L CRP
    5 mg/L CRP
    53 mg/L CRP
    PreaquisitionDiagnosis/ patient condition
    Hospitalized with inflammatory disease
    Biospecimen AcquisitionType of collection container/solution
    Glass tube
    Serum separator tube
    Biospecimen Aliquots and ComponentsAliquot size/volume
    0.5 mL
    1 mL
    1.5 mL
    2 mL
    StorageTime at room temperature
    0 min
    15 min