Falsely increased free triiodothyronine in sera stored in serum separator tubes.
Kilinç AS, Düzoylum A, Uncugil CF, Yücel D
Purpose
Conclusion
Studies (1)
Purpose
The purpose of this study was to determine the effects of refrigerated storage of serum in SST or glass collection tubes on free T3, free T4, and TSH.
Summary of Findings
Collection container and storage at 4 degrees C did not affect free T4 or TSH levels in serum. Free T3 was higher when serum was collected in SST than when serum was collected in plain glass tubes or vacuettes. Storage of serum with separator gel, at 4 degrees C, for up to 12 h did not affect free T3. However, when serum was stored in vacutainer SST or glass tubes for more than 24 at 4 degrees C, there was a significant effect of storage duration on T3.
Biospecimens
FluidSerumFluidBloodPreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
PeptideClinical chemistry/auto analyzerGlycoproteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution GlassVacuette SSTVacutainer SSTStorage Storage duration 12 h24 h4 h40 min48 h72 h8 hStorage Storage temperature 4 degrees CStorage Type of storage container Plain tubeSST with separator gel